standard wavelength used for correction should also be from the

atomic state, such as Sc 361.383. Conversely, match ionic sample

lines with ionic internal standard lines. (

Note

: Do not use yttrium as

an internal standard, since it is found native at low levels in some

phosphate ore sources.)

(

b

)

ICP wavelengths.

—A number of wavelengths may be used

for analysis of the 8 elements of interest, depending on the capability

of the analytical instrument used. As a minimum, select at least 2

wavelengths for each element of interest, and report the average

value of closely agreeing results, except for lead and selenium, for

which there is only one reliable wavelength available. Following is a

list of suggested wavelengths, not in any priority order, that have

been found acceptable for most fertilizer materials. Other lines of

appropriate sensitivity, free of interferences or corrected for

interferences, may be just as acceptable. However, it is imperative

that instrument response (both instrument graphic output and

calculated concentration) be reviewed for each sample and element.

Fertilizer materials are extremely variable in composition, and a

wide concentration range of potential interfering elements is

expected, so no single wavelength will work in every instance.

Occasional data with interference will inevitably be found, and must

be eliminated from inclusion in the mean calculation for that

particular element and sample.

Wavelengths (nm): As: 188.980, 193.696; Cd: 214.439, 226.502;

Co: 228.615, 230.786, 235.341; Cr: 205.560, 267.716, 276.653;

Mo: 201.512, 202.032, 203.846, 204.598; Ni: 216.555, 222.295,

222.486, 227.877, 231.604, 239.452; Pb: 220.353; Se: 196.026; Sc:

361.383, 431.408; Be: 234.861, 249.473.

(

c

)

Wavelength interference treatment.

—Interelement

interference can cause substantial error in analytical result. Error

can be minimized by several techniques: (

1

) Three or more

analytical lines may be used for a given element, and when an

interferent is present in a particular line, the result for that line is

omitted from the mean value reported. (

2

) Certain vendors’

instrument software has the capability of mathematically modeling

potential interferents, and deconvoluting the instrument response

into an analytical element portion and an interferent portion. (

3

)

Interelement correction is an alternative mathematical technique to

use with instruments for which mathematic modeling is not

available, or where direct spectral overlap negates use of the

deconvolution technique.

The following lines, if used, must utilize one of the correction

techniques; corrections for other lines may be applied as needed and

appropriate: (

1

) As 188.980: Correct for Cr interference at 188.995,

or verify that Cr is not present in the test portion analyzed. (

2

) As

193.696: Fe affects the arsenic peak. Remove with an Fe model, or

verify that Fe is not present in the test portion analyzed. (

3

) Cd

214.439 and 226.502: Fe, present in many fertilizers, interferes with

both suggested Cd wavelengths. Mathematically correct instrument

Cd response for the interference, or verify analytically that Fe is not

present in the test portion analyzed. (

4

) Pb 220.353: Mathematically

correct instrument Pb response for Fe interference, or verify that Fe

is not present in the test portion analyzed. (

5

) Se 196.026:

Mathematically correct instrument Se response for Fe interference,

or verify that Fe is not present in the test portion analyzed.

(

d

)

ICP instrument calibration

.—Prepare working standard

solutions from commercial stock standards at 1000 mg/kg. Custom

blended multielement stock standard in HNO

3

is acceptable. Prepare a

minimum of 5 working standards at 0.1, 0.5, 1.0, 5.0, and 10.0 mg/kg,

plus blank, of each element, in 10% trace metal grade HNO

3

. Working

standards should be in the linear range, with correlation coefficients

of at least 0.9999.

G. Reagents

(

a

)

Water

.—Use 18 Megaohm water for dilution.

(

b

)

HNO

3

.—Use trace metal grade HNO

3

.

(

c

)

0.5% Triton X100 solution

.—Dilute 0.5 mL Triton X100 to

100 mL with H

2

O.

(

d

)

Ionization buffer/internal standard solution.—

Weigh 8.0 g

CsCl into a 1000 mL acid-washed volumetric flask. Add 3 mL each

of ICP grade scandium and beryllium 1000 mg/kg stock solution, as

internal standards. Also add 1 mL of 0.5% Triton X100, dilute to

volume, and mix. Store in a polypropylene bottle. (

Note

: Reagent

concentrations assume the use of white/white, 1.02 mm id sample

pump tube, and orange/white, 0.64 mm id internal standard pump

tube. If the sample and internal standard solutions are mixed in

different proportions by the instrument’s peristaltic pump, then

adjust the reagent concentrations to meet concentration

requirements of mixed solution nebulized by the instrument, as

outlined in

F

. Note that sample and internal standard solutionmixing

ratio is proportional to pump tube flow rates, not proportional to

pump tube IDs.)

(

e

)

Stock standard solution.

—Working standards can be

prepared from ICP grade individual element 1000 mg/kg

commercial stock standard solutions. However, it is also acceptable

to use commercially prepared custom blended stock standard

mixtures containing all of the 8 elements at 1000 mg/kg. A number

of companies provide this stock standard service.

(

f

)

10 mg/kg

intermediate stock standard solution for

preparation of low-level working standards.

—Dilute 5.0 mL of

stock standard solution to 500mL. Prepare fresh each time standards

are prepared, and use immediately after preparation.

(

g

)

Working standard solutions.

—Standards are designed to have

the same acid concentration as digested test solutions. Date all

calibration solutions when made, which should be stable for at least

1 month, but not longer than 2 months. Monitor standard curve fit

and intensity for signs of change and degradation over time. (

1

)

10 mg/kg elements

.—Pipet 5.0 mL of combined 1000 mg/kg

element stock solution into a 500 mL acid-washed volumetric flask.

Add 50 mL of trace metal grade HNO

3

, dilute to volume with H

2

O,

mix, and transfer to acid-washed polypropylene bottle. (

2

)

5 mg/kg

elements

.—Pipet 5.0 mL of combined 1000 mg/kg element stock

solution into a 1000 mL acid-washed volumetric flask. Add 100 mL

of trace metal grade HNO

3

, dilute to volume with H

2

O, mix, and

transfer to an acid-washed polypropylene bottle. (

3

)

1 mg/kg

elements

.—Pipet 50.0 mL of 10 mg/kg intermediate stock solution

into a 500 mL acid-washed volumetric flask. Add 50 mL of trace

metal grade HNO

3

, dilute to volume with H

2

O, mix, and transfer to

an acid-washed polypropy lene bott le. (

4

)

0.5 mg/kg

elements

.—Pipet 25.0 mL of 10 mg/kg intermediate stock solution

into a 500 mL acid-washed volumetric flask. Add 50 mL of trace

metal grade HNO

3

, dilute to volume with H

2

O, mix, and transfer to

an acid-washed polypropy lene bott le. (

5

)

0.1 mg/kg

elements

.—Pipet 5.0 mL of 10 mg/kg intermediate stock solution

into a 500 mL acid-washed volumetric flask. Add 50 mL of trace

metal grade HNO

3

, dilute to volume with H

2

O, mix, and transfer to

an acid-washed polypropylene bottle. (

6

)

0.0 mg/kg elements

(blank)

.—Add 50 mL of trace metal grade HNO

3

into a 500 mL