Table of Contents Table of Contents
Previous Page  117 / 172 Next Page
Information
Show Menu
Previous Page 117 / 172 Next Page
Page Background

volumetric flask, dilute to volume with H

2

O, mix, and transfer to an

acid-washed polypropylene bottle.

(

h

)

Sampler wash solution, 1% HNO

3

.—Dilute 10 mL of trace

metal grade HNO

3

to 1000 mL with H

2

O.

H. Determination

Analyze test solutions using an ICP-OES instrument calibrated

with standard solutions. Insert a 10 mg/kg working standard or other

suitable quality control solution every 10 test portions, to monitor

for instrument drift. The inclusion of a digestion blank, a sample

duplicate, and known reference materials is highly encouraged.

I. Calculations

Element, mg / kg = (instrument concentration, mg / L)(100 mL)(1 L / 1000 mL)

(sample weight, g)(1 kg / 1000 g)

where 100mL assumes themicrowave digest is diluted to 100mL.

Reference:

J. AOAC Int. 89, 1447(2006)

.