NCS Report on VMAT

Workload can be reduced by using class solutions for each tumour site

A class solution should consist of…

the required structures to be used in optimisation and the allowed ranges of

the associated optimisation objectives, constraints and weights

a standard set of technical requirements :

the number of fields

desired start and stop angles

collimator settings

beam energies

control point spacing

dose grid size

By controlling the range of parameters using class solutions it promotes