Detector

Appropiate for

Be careful

Ionization

chambers

For x-ray depth dose

measurements

Plane-parallel chamber for

build-up region

Output factors large fields

Output factors for small fields

(micro-chambers)

Check polarity effect for both

photon and electron beams

Always measure with the same

nominal V

For micro-chambers the stem

signal can be an issue

Correct by Sw,air for electron

depth dose measurements.

Use small volume detectors for

scanning

Effective point of measurement

Check stability temperature

during measurements

Diodes

Depth dose measurements

and profiles

Shielded detector for x-rays

Non-shielded for electron

beams

Output factors

Shielded detector

standard/large fields

Non-shielded for small fields

Energy dependence

Sensitivity variation with

accumulated dose

May present sensitivity variation

with dose rate

Check stability temperature

during measurements