Oncentra/Monaco – Pencil beam and Collapsed
Cone Convolution
❑
Patient tissue distribution is represented by a 3D density matrix where the
properties in each voxel are either derived from pixel values in a CT image
or from user specified values of mass or electron density.
❑
It is assumed that the specified density values correspond to tissue material
of “standard” compositions (see ICRP 23 and ICRU 44).
❑
The Hounsfield number assigned to each standard tissue is mapped from
the calculated electron density using relations given by Knöös et al [38].
❑
…for charged particle equilibrium conditions CC calculates dose to medium
while PK calculates dose to a large water cavity in medium. Both algorithms
consider attenuation of the beam as per medium
thus;
❑
Pencil kernel and for point kernel
Dublin 2018
15
,
,
Previously TMS-Helax, Oncentra MasterPlan