Conclusions

Dublin 2018

32

The results from this study show that contrary to the popular

perception, conventional photon dose calculation using water with

relative electron densities produces dose distributions that are much

closer to Monte Carlo calculated dose-to-medium distributions than to

dose-to-water distributions converted using mass stopping-power ratios

of water to bone…

… explained by the fact that the photon dose deposition depends

mainly on the electron density since the predominant mode of

interaction is Compton scattering for photon beams of radiotherapy

interest.

Thus, the dose deposition scaled based on the relative electron

density is effectively dose to medium

(with the corresponding

electron density), as implemented in many commercial treatment

planning dose algorithms in spite of their initial intentions or claimed

applications.

Ma and Li, 2011