The following data are provided:
1. 6 MV Photon beam with a quality index of 0.676
2. A Farmer chamber (PTW 30013) was used
3. Electrometer reading: Charge is measured 5 times: 9.84, 9.99,
10.04, 9.98, and 10.06 [nC]
4. Background level: 0.001
±
0.001 [nC]
5. Calibration factor: 5.772 10
7
[Gy(C]; U = 2.2%
The compilation of all input uncertainties and associated sensitivity
factors with subsequent evaluation is referred to as establishing
an
Uncertainty budget