Should IMRT patient QC be done field-by-field or

for the composite plan?

Assume a 5 field plan each beam contributes with 1/5 to

the target

Each beam are delivered within 3 % of calculation

Except for one that is 15% off.

This

plan

will not pass field-by-field QC

For a composite plan you

will not

detect this, the plan is

within 3%!!!