EPID CALIBRATION
Correction factors (at this moment to be used for pretreat verifications
and linac QA)
DEPENDING ON THE DETECTOR: DIFFERENT CORRECTION FACTORS CAN BE EXPECTED
❑
Field size
❑
Backscatter from the EPID arm (Varian aS500, aS1000).
Not needed for aS1200 (backscatter shielding added)
❑
EPID SAG if used at different gantry angles
THE INTENSITY AT EACH POINT IN THE IMAGE MUST BE CALIBRATED IN A
QUANTITATIVE SENSE
❑
The dose response of each pixel point must be known
❑
The effects of scattered radiation in the detector must be understood
❑
The temporal stability of the detector must be known.