Forward Planning / Optimization

Dose Points-based Adjustment of Dwell Times T

i

:

Graphical Optimisation

Planner takes an isodose line and shifts it to another

position:

-

Take a line: Starting Point defines the final Dose Level, D

final

-

Shift to another position: This point becomes the DP of the

“optimisation” and has at the time of planner action the dose D

start

.

-

The Dwell Times T

i

have to be adjusted in such a way, that our DP

gets instead of D

start

the dose level D

final

.

The Dwell Times (all or part of them) are then adjusted according to:

start

final

D -

D D