Table of Contents Table of Contents
Previous Page  145 / 1082 Next Page
Information
Show Menu
Previous Page 145 / 1082 Next Page
Page Background

S132

ESTRO 36 2017

_______________________________________________________________________________________________

Conclusion

These results show that T1 and T2 mapping, DCE and DWI

are robust over different imaging platforms. The quality

of T1 and T2 mapping on the MR-Linac has been

demonstrated using a scan of a healthy volunteer.

OC-0260 Experimental verification of dose

enhancement effects in a lung phantom from inline

mag netic fields

B.M. Oborn

1

, M. Gargett

2

, T. Causer

1

, N. Hardcastle

2

, P.

Metcalfe

3

, P. Keall

4

1

Illawarra Cancer Care Centre, Medical Physics,

Wollongong NSW, Australia

2

Royal North Shore Hospital, Medical Physics, Sydney,

Australia

3

University of Wollongong, Centre for Medical Radiation

Physics, Wollongong, Australia

4

Universityof Sydney, School of Medicine, Sydney,

Australia

Purpose or Objective

To present for the first time experimental evidence of lung

dose enhancement effects caused by stro ng inline

magnetic fields. Such fields will be typical of future inline

MRI-linac systems.

Material and Methods

A permanent magnet device was utilized to generate a 1.2

T magnetic field that encompassed a small lung-

equivalent phantom of 0.3 g/cm

3

. 6MV and 10 MV photon

beams with 0.5 x 3 cm

2

field size were incident parallel

with the magnetic field direction and EBT3 film was placed

inside the lung phantom (Fig 1 shows a schematic setup).

Monte Carlo simulations were also performed of the

identical setup, and of larger field sizes.

Figure 1. Schematic diagram of the lung phantom setup

and magnetic field orientation.

Results

Inclusion of the 1.2 T inline magnetic field induced a 12%

(6MV) and 14% (10MV) increase in the dose along the beam

central axis as compared to the 0 T reference case (Fig 2).

This increase arises from a relative reduction in the

amount of lateral secondary electron scatter in the low

density lung medium. Monte Carlo modeling matched well

(+/- 2%) to the experimentally observed results. For

simulated larger field sizes the enhancement effect drops

off to around 3% at 5x5 cm

2

, which is a natural result of

charged particle equilibrium being restored at larger field

sizes. This dose enhancement effect will be expected to

compliment the precise treatment of small tumors

embedded in lung tissue using real-time inline MRI-linac

systems.

Figure 2. Film and profile results for the 6MV beam.