Table of Contents Table of Contents
Previous Page  490 / 1082 Next Page
Information
Show Menu
Previous Page 490 / 1082 Next Page
Page Background

S475

ESTRO 36 2017

_______________________________________________________________________________________________

protons, IMPT

PR

plans were the most robust and only 4/26

(15%) decreased in coverage to below 99%. The CTV

coverage for all patients and plans are shown in Fig 1. The

most common anatomical changes are lateral target

deformations, enlargement of mediastinum and changes

in diaphragm position. The posterior proton plans are

sensitive to target deformations, while the multiple

photon fields are sensitive to all three types of changes

(see

Fig.

2).

Conclusion

Treating oesophageal cancer with protons has the

advantage of decreasing dose to organs at risk and at the

same time it improves the robustness towards common

anatomical changes. Frequent imaging is still needed to

identify patients with target deformations requiring

adaptive treatment planning.

PO-0878 Plan adaptation on the MR-Linac: first

dosimetric validation of a simple dose shift

R. Koopman

1

, A.J.A.J. Van de Schoot

1

, J. Kaas

1

, T. Perik

1

,

T.M. Janssen

1

, U.A. Van der Heide

1

, J.J. Sonke

1

1

Netherlands Cancer Institute Antoni van Leeuwenhoek

Hospital, Radiation Oncology, Amsterdam, The

Netherlands

Purpose or Objective

Patient positioning on the MR-Linac (MRL; Elekta AB,

Stockholm) requires online plan adaptations to correct for

setup errors due to the fixed couch position. The aim of

our study was to validate the size and direction of such

plan adaptations (simple dose shifts) and evaluate

dosimetric differences for rectum cancer patients.

Material and Methods

The planning CT and delineated structures of four rectum

cancer patients were selected. For each patient, a MRL

treatment plan was generated with Monaco using a 7-

beam IMRT technique (25 x 2.0 Gy) including all MRL-

specific properties (7MV, 1.5 T magnetic field, collimator

90°, FFF, SAD: 143.5 cm). Patient setup errors of 1.0 cm

and 2.0 cm in the CC and LR directions were simulated by

shifting the planning CT with respect to the isocenter

position. For each setup error, the initial plan was adapted

by first adjusting the leaves of each segment to

approximate the shift and second re-optimize the weight

of each segment. Also, a reference plan was generated by

adapting the initial plan with a 0.0 cm shift, as the second

phase of plan adaptation was observed to introduce dose

changes even for a 0.0 cm shift. All plans were rescaled

(PTV V

95%

= 99%). The reference and adapted plans were

irradiated on the MRL on a slab phantom with a 2D

detector array (PTW Octavius 1500

MR

) inserted parallel to

the couch at the center position of the PTV. For each plan,

the phantom position was changed according to the

introduced shift. Patient setup errors in the AP direction

cannot be evaluated using this measurement setup. The

measured 2D dose distribution of the reference plan was

rigidly registered to the measured 2D dose distribution of

the adapted plans in order to assess the positional

accuracy of the simple dose shift. After alignment, the

similarity between the 2D dose distributions of the

reference plan and the adapted plans was evaluated using

a 3%/3mm γ analysis (local dose, 20% low dose threshold).

Results

For all adapted plans, the measured positional accuracy

was within 0.1 cm. The γ analysis between the dose

distributions of the reference plan and the adapted plans

resulted in an average pass rate (γ≤1) of

96.2% (range: 83.3% – 99.9%). Smaller values of γ

mean

were

observed for dose shifts in the CC direction compared to

the LR direction as well as for 1.0 cm dose shifts compared

to 2.0 cm dose shifts (Table 1). Figure 1 shows an example

of a 2D γ distribution. High γ values are measured in the

low dose area mainly. A simple dose shift to correct for a

1.0 cm setup error in the CC direction resulted in limited

dose differences. Various γ hotspots were observed for

the 2.0 cm setup error in the LR direction.