S502
ESTRO 36 2017
_______________________________________________________________________________________________
3
Czech Technical University in Prague, Institute of
Experimental and Applied Physics, Prague, Czech
Republic
4
German Cancer Research Center DKFZ, Department of
Medical Physics, Heidelberg, Germany
Purpose or Objective
(1) Purpose: In highly conformal radiotherapy, like ion
beam radiotherapy, inter- and intrafractional monitoring
of the target is desirable. Due to the steepness of the
rising part of the Bragg curve, ion beam radiography can
in principle provide high resolution of the traversed tissue
thickness. Ion beam radiography is furthermore attractive
due to its potential to measure the stopping power of the
tissue directly. However, currently there is no detection
system for clinical imaging of patients. Helium ions as the
imaging modality provide the advantages of decreased
multiple scattering in comparison to protons and lower
biological effectiveness than the carbon ions.
Material and Methods
(2) Methods: Plastic phantoms contain ning 1mm deep
step-like inhomogeneities were imaged with helium ion
beams at the Heidelberg Ion Beam Therapy facility in
Germany. To register the radiation, a system of 5 parallel
layers, based on the semiconductor pixelized detector
Timepix, which was developed by the Medipix
Collaboration at CERN, was used. Two layers in front of
the phantom enabled us to measure the position and
direction of incoming helium ions. Another pair of
detection layers, located behind the phantom, registered
the outgoing particles and an additional layer was used to
measure their energy loss and to identify the ion type.
Synchronization of all the five detector layers enabled us
to associate the outgoing particles to the incoming ones.
To build the image of the phantom, we used the measured
information about the transversal position of the incoming
and outgoing particle, their direction and type (He or H).
Results
(3) Results: With this system we imaged a 1 mm step in a
160 mm thick PMMA phantom. Spatial resolution below 2
mm was reached when the inhomogeneity was located in
the phantom, while resolution below 1 mm was achieved
in the cases where the step was located at the front or at
the end of the phantom. Hereby we have shown that the
information about flight direction of the incoming and
outgoing ion, together with the capability to identify them
and thus select solely helium ions, enables to improve the
spatial resolution by a factor of more than three.
Conclusion
(4) Conclusion: We have shown experimentally that helium
beam radiography reaches in simple phantoms spatial
resolution in the region which is attractive for highly
conformal radiotherapy. In the presentation the results
obtained with helium beams as the imaging modality will
be compared to proton-based imaging.
PO-0915 Performance study of a prototype straight-
through linac delivery system with an EPID assembly
R. Scheuermann
1
, C. Kennedy
1
, D. Mihailidis
1
, J. Metz
1
1
University of Pennsylvania, Radiation Oncology,
Philadelphia, USA
Purpose or Objective
To study and expand the use of the Machine Performance
Check (MPC) tool in monitoring the continuous operational
performance of a prototype delivery system composed of
a straight-through-linac and an in-line MV portal imaging
panel (Proof-of-Concept).
Material and Methods
The MPC, as implemented in TrueBeam (TB), is an
integrated self-check tool that assures that critical
machine performance are within specifications, e.g.
mechanical accuracy and radiation output. As adapted to
the prototype straight-through linac delivering 6X-FFF
(filter-free) beam, the automated tests are based on its
in-line MV EPID. The MPC acquires a series of MV images
of an IsoCal phantom, capturing beam properties and
mechanical data such as MLC and gantry accuracies. A
new MPC test monitors output stability in terms of percent
deviation from the baseline data of the actual measured
beam. All measured data are automatically processed,
analyzed, and displayed for evaluation, thus providing a
reliable and fast method for routine machine performance
assessment. Independent tests such as star-shots,
Winston-Lutz, MLC picket fence patterns and output
measurements on a daily basis were employed to
benchmark the MPC test results for the prototype system.
Results
MPC results were collected daily for six months on both
the prototype and a TB. The independent tests on the
prototype system were repeated weekly to validate the
MPC results. A sample comparison of the MPC results for
the prototype against independent tests are shown in
Table 1. The output stability of the prototype system, as
measured with the MPC and a DailyQA
TM
3 device, is
comparable (Fig. 1), and within 0.5% of independent
output measurement for the period shown. All tests
performed were within the tolerances allowed by the MPC
and agreed in most cases with the result of the
independent tests. The prototype system performs as well
as the TB system. A summary of MPC test results and
comparisons with independent measurements will be
shown alongside with the TB MPC results.
Table 1
Figure 1
Conclusion
The performance of a straight-through linac d elivering 6X-
FFF (filter-free) beam with an EPID panel wa s investigated
with the MPC testing tool and that method w as validated
against independent tests for proof of con cept. MPC is a
complete, reliable and quick test suite that monitors the
performance of a treatment unit on routine basis.