S818
ESTRO 36 2017
_______________________________________________________________________________________________
CT volumes at arbitrary positions within the respiratory
cycle. Daily CBCT images are used to calculate a baseline
shift between the planning CT and the actual anatomy
using deformable image registration. Treatment planning
was done by the TPS Pinnacle³ v9.8 (Phillips Radiation
Oncology Systems, Fitchburg, WI, USA). The motion model
was used to estimate CT volumes in intervals of 100 ms
that were imported into the TPS to calculate a partial
delivered dose. The Vero system uses rotations of the
LINAC head for DT, so the resulting images have to be
rotated according to the pan/tilt motion since the TPS is
not able to model a non-axial beam line. The partial doses
were back-rotated and superposed resulting in the
actually delivered 4D dose distribution. In addition, 4D
dose-volume-histograms (
DVH
) were calculated by
warping the reference contours onto every respiratory
phase using the motion model. The resulting 4D dose
distributions were evaluated and compared to phantom
measurements using the ArcCheck (Sun Nuclear,
Melbourne, FL, USA) in combination with a motion
platform jointly developed with QRM, Möhrendorf,
Germany.
Results
The accuracy of the motion model was validated for 20
patients with a geometrical uncertainty of < 3 mm. Dose
comparison of the estimated volumes to clinical ground-
truth CTs resulted in a pass rate above 98.9% for a γ-
criterion of 2% / 2 mm. 4D dose distributions could be
reconstructed and were in good agreement to phantom
measurements. The actually delivered dose for two liver
patients was recalculated and additional patients are
currently ongoing.
Conclusion
The presented approach is feasible to reconstruct 4D dose
distributions for DT patients using a common TPS together
with the presented motion model. External surrogates to
calculate the breathing state are essential as well as daily
CBCT or kV images to correct for baseline shifts.
EP-1542 Can proton therapy for head and neck cancer
reduce side effects while maintaining target
robustness?
D. Scandurra
1
, R.G.J. Kierkels
1
, M. Gelderman
1
, H.M.
Credoe
1
, H.P. Bijl
1
, R.J.H.M. Steenbakkers
1
, J.G.M.
Vemer - van de Hoek
1
, J.A. Langendijk
1
1
University Medical Center Groningen, Department of
Radiation Oncology, Groningen, The Netherlands
Purpose or Objective
Head and neck cancer generally consists of targets
requiring high doses of radiation in close proximity to vital
healthy organs. Proton therapy leads to a reduction of
integral dose and therefore can reduce normal tissue
complication probabilities (NTCP). The proton range,
however, depends on the material along its path and is
therefore susceptible to uncertainties, potentially
degrading target coverage. These uncertainties can be
partially accounted for by conventional target margins
(PTV) or robust-optimisation approaches. In this in-silico
study, four variations of pencil beam scanning (PBS)
proton therapy plans were designed for each patient and
compared to the clinical photon plan in terms of
robustness and NTCP reduction.
Material and Methods
Four PBS plans were made for each patient using
RayStation (v4.99 RaySearch Laboratories AB, Sweden):
1) PTV based, multi-field optimisation (PTV-MFO),
2) PTV based, single field optimisation (PTV-SFO),
3) CTV robustly optimised MFO (CTV-MFO), and
4) CTV robustly optimised SFO (CTV-SFO).
Each plan was designed to a similar target homogeneity
index and normalised to CTV70 D98%. Four beams were
used in an ‘x’ arrangement and a 4 cm range shifter was
used where appropriate.The PTV was defined as 5 mm
geometric expansion to CTV, identical to current photon
plans in our clinic (VMAT). Robust optimisation settings
were 5 mm isotropically and ±3% HU uncertainty. For each
plan, robust evaluation was performed for a combination
of ±5 mm translational, ±2
⁰
rotational and ±3% HU errors
(>50 scenarios per plan). The error scenario with the
lowest target coverage (V95% to CTV70) is deemed the
‘worst-case’
scenario.
Organ at risk (OAR) doses and NTCP values of the nominal
proton plans and the clinical VMAT plan were compared.
Results
Robust evaluation showed CTV70 target coverage
degraded significantly for PTV based plans, particularly
using MFO (-15%) but also for SFO (-5.4%). Robustly
optimised plans performed much better, particularly CTV-
MFO (-1.7%) which comes very close to matching the
robustness of the clinical VMAT plan (-1.2%) (figure 1).
CTV-SFO under-performed due to the beam arrangement
and the high weighting on OAR dose reduction.
OAR doses were significantly better in MFO plans, as this
gave the optimiser the most freedom in per–beam dose
distribution. In particular, CTV-MFO plans had the lowest
OAR doses of all proton and photon plans. NTCP model
calculations show that, compared to our clinical VMAT
plans, reductions across all patients for xerostomia (-
5.9%), dysphagia (-7.0%) and tube feeding (-4.7%) can be
expected, but these values varied widely among different
tumour sites, stages and individual patients (table 1).
Conclusion
Compared to photon VMAT plans, robustly optimised PBS
proton therapy plans reduce OAR doses in head and neck
cancer while maintaining a high degree of target coverage
robustness. NTCP calculations show a clinical benefit can
be expected for a significant proportion of patients.
EP-1543 Dominant intraprostatic lesions boosting:
comparison of tomotherapy, VMAT and IMPT
P. Andrzejewski
1
, A. Jodda
2
, P. Kuess
1
, D. Georg
1
, J.
Malicki
3
, T. Piotrowski
3
1
Medical University Vienna, Dept. of Radiation Oncology-
Christian Doppler Laboratory for Medical Radiation
Research for Radiation Oncology, Vienna, Austria
2
Greater Poland Cancer Centre, Department of Medical
Physics, Poznan, Poland