2b. QC of the PET system
• Periodic measurements recommended for
-
Transaxial resolution
-
Imaging Scale
-
Documentation unit
• Routine PET QC measurements
-
Buseman Sokole et al, EJNMMI 2010; 37: 662-71.
-
Buseman Sokole et al, EJNMMI 2010; 37: 672-81.
• QC of PET components consists of
-
Check of coincidence sensitivity and detector normalization
-
Normalization calibration (conversion factor for activity conc.)
-
Certification of imaging properties that describes the results of
mandatory data correction (e.g. randoms, attenuation, scatter).
10 | D. Thorwarth | 09/2015
ESTRO teaching course on advanced imaging techniques