2. Quality control, calibration
13 | D. Thorwarth | 09/2015
ESTRO teaching course on advanced imaging techniques
Recommendations
• Regular QC and calibration of PET/CT hardware is required for
PET/CT-guided RT TP. Existing guidelines and recommendations
should be followed.
• There is a lack of standard QC procedures for ancillary RT
devices. A set of QC measured should be agreed on among expert
staff from nuclear medicine, radiology and radiation oncology.
• The full range of service operations must be verified following
service access of PET/CT hardware