Table of Contents Table of Contents
Previous Page  95 / 156 Next Page
Information
Show Menu
Previous Page 95 / 156 Next Page
Page Background

Additional adaptations

SMPR 2011.006

OLD

Limit of detection (LOD)

.—The minimum concentration or mass of

analyte that can be detected in a given matrix with no greater than

5% false-positive risk and 5% false-negative risk.

NEW

Limit of Detection (LOD)

The smallest amount or concentration of an analyte that can be

estimated with acceptable reliability. Determined as: LOD = blank

mean + 3 standard deviations of ten independent analyses of blank or

blank spiked at low level (to be agreed upon by Study Directors) (if

there is no detectable blank signal). See reference to Appendix L in

section 7. Validation Guidance.

5

Additional adaptations

SMPR 2011.006

6 Reference Material(s)

National Institute of Standards and Technology (NIST) Standard Reference

Material (SRM) 1849

a

Infant/Adult Nutritional Formula, or equivalent. The SRM

is a milk-based, hybrid infant/adult nutritional powder prepared by a

manufacturer of infant formula and adult nutritional products. A unit of SRM

1849

a

consists of 10 packets, each containing approximately 10 g of

material. Certified value of folic acid in NIST 1849 is 2.11 (±0.13) mg/kg.

Note: The reference value for NIST 1849 is defined in terms of folic acid. The

performance parameters in this SMPR are intended for folate and 5-methyl-

tetrahydrofolate polyglutamate. Some discrepancy may be expected.

6