Example:     AOAC Official Method 2015.01 

Heavy Metals in Food 

Inductively Coupled Plasma–Mass Spectrometry 

First Action 2015 

Note

: The following is not intended to be used as a comprehensive training manual. Analytical procedures are 

written based on the assumption that they will be performed by technicians who are formally trained in at least 

the basic principles of chemical analysis and in the use of the subject technology. 

{Applicable for the determination of heavy metals [arsenic (As), CAS No. 7440‐38‐2; cadmium (Cd), CAS No. 7440‐

43‐9; lead (Pb), CAS No. 7439‐92‐1; and mercury (Hg), CAS No. 7439‐97‐6] at trace levels in food and beverage 

samples, including solid chocolate, fruit juice, fish, infant formula, and rice, using microwave digestion and 

inductively coupled plasma–mass spectrometry (ICP‐MS).} 

Caution

: Nitric acid and hydrochloric acid are corrosive. When working with these acids, wear adequate protective 

gear, including eye protection, gloves with the appropriate resistance, and a laboratory coat. Use an adequate 

fume hood for all acids. Hydrogen peroxide is a strong oxidizer and can react violently with organic material to give 

off oxygen gas and heat. Adequate protective gear should be worn. Many of the chemicals have toxicities that are 

not well established and must be handled with care. For all known chemicals used, consult the Material Safety 

Data Sheet (MSDS) in advance.  Etc., etc., etc. 

A. Method Performance 

The performance criteria in Table 2015.01A is  prescribed by 

Standard Methods Performance Requirements

 (SMPR) 

2012.00

7

, which was adopted by the AOAC Stakeholder Panel  for Dietary Supplements  (SPDS).  Laboratory(ies) 

implementing this method should follow ISO Standard 17025 to verify that the performance of this method in your 

laboratory meets the criteria in Table 2015.11A.    Please report verification results to AOAC at this link:  

AOAC 

2015.01

Table 2015.01A Verification Acceptance Criteria 

Limit of Quantitation (LOQ) 

Parameters 

Ranges 

Repeatability (RSD

r

Recovery 

≥ 10 ppb to 100 ppb 

≤ 15% 

60% ‐ 115% 

>100 ppb to 1 ppm 

 ≤ 11% 

80% ‐ 115% 

> 1 ppm to 10 ppm 

≤ 7.3% 

80% ‐ 115% 

B. Principle 

Food samples are thoroughly homogenized and then prepared by microwave digestion and the addition of dilute 

solutions of gold (Au) and lutetium (Lu). The Au is used to stabilize the Hg in the preparation, and the Lu is used to 

AOAC OMB Meeting

February 10-11, 2016