Table of Contents Table of Contents
Previous Page  119 / 670 Next Page
Information
Show Menu
Previous Page 119 / 670 Next Page
Page Background

Table 5. Suitable range of sample concentration after dilution.

SRM

Product

Code

Reference

value

Sample

size

HPLC-

Injected

Sample

Measured

value

Recovery

vs CoA

Calibration

Standard

Suitable

range

(mg/kg

powder)

(g)

Conc.

(ng/mL)

(mg/kg

powder)

(%)

Level

Conc.

(ng/mL)

Conc.

(ng/mL)

1849a

CLC-10b

1.99

0.1445

6.276

2.172

109

Blank

0

8.5 – 60

0.1532

6.981

2.278

114

1

4.981

0.2006

8.454

2.107

106

2

9.962

1.0020

40.21

2.006

101

3

19.92

1.0006

41.05

2.051

103

4

39.85

1.5177

58.69

1.934

97.2

5

59.77

1.5082

57.60

1.910

96.0

6

79.70

2.0006

69.39

1.734

87.1

7

99.62

2.0003

69.28

1.732

87.0

Table 4. A representative calibration standard data.

Level

Concentration

(ng/mL)

HPLC Peak area

Amount

(ng/mL)

Error

1

5

0.2019

6.21

24%

2

10

0.6600

9.00

10%

3

20

2.2932

18.61

7%

4

40

6.1743

41.03

3%

5

60

9.2305

60.45

1%

6

80

11.353

77.66

3%

7

100

12.863

103.9

4%

Figure 1. A representative calibration curve.

Bio-03 (February 2016)

FOR ERP USE ONLY

DO NOT DISTRIBUTE