Quantitative Analysis Summary Report
Sample Type
Response
Final Conc
Exp Conc
Accuracy
Calibration
37598
10.1260
10.0000
101.26
Calibration
71576
21.2075
20.0000
106.04
Calibration
167953
52.6403
50.0000
105.28
Calibration
317424
101.3894
100.0000
101.39
Calibration
604434
194.9961
200.0000
97.50
Calibration
831510
269.0554
300.0000
89.69
Calibration
1000088
324.0361
400.0000
81.01
Sample
6271
0.0000
Sample
39001
10.5835
Sample
40054
10.9267
Sample
39267
10.6702
Sample
39344
10.6954
Sample
39896
10.8752
Sample
39580
10.7723
Target Compound
Riboflavin
Sample Type
Response
Final Conc
Exp Conc
Accuracy
Calibration
1200
2.0272
1.0000
202.72
Calibration
1883
2.8130
2.0000
140.65
Calibration
4348
5.6511
5.0000
113.02
Calibration
7749
9.5651
10.0000
95.65
Calibration
16473
19.6067
20.0000
98.03
Calibration
40826
47.6370
50.0000
95.27
Calibration
82575
95.6892
100.0000
95.69
Calibration
173711
200.5872
200.0000
100.29
Calibration
262535
302.8232
300.0000
100.94
Calibration
354048
408.1546
400.0000
102.04
Sample
531
1.2572
Sample
88826
102.8850
Sample
86761
100.5081
Sample
87018
100.8037
Sample
87653
101.5343
Sample
86242
99.9109
Sample
87028
100.8153
01-01-2015-WSV-VAL-P&A-1.xlsx
Page 4 of 77
TUV-SUD South Asia, Bangalore
Printed at: 4:09 PM on: 2/8/2016