peak noise between 6.0 and 7.0 min over five injections of a prepared sample. The estimated
LOD and LOQ for fluoride were 1.27 and 4.24
μ
g/100g respectively. The estimated LOD and
LOQ for chloride were 0.006 and 0.02
m
g/100g respectively.
Summary
The accuracy data for fluoride and chloride met the criteria specified in the SMPR. The
limit of quantitation for fluoride was estimated to be 4.24
μ
g/100g. The limit of detection was
estimated to be 1.27
μ
g/100g. Acceptable linearity was demonstrated for both fluoride and
chloride over a range specified in SMPR.
This method is found to be suitable for the analysis of fluoride and chloride in a single
method for all SPIFAN matrices.
Fluor-02/Chlor-03 (February 2016)