Protective shielding
Collimator channels
Leksell
®
Coordinate
Frame
Patient
positioning system
Radiation sources
Isocenter/
Target in the brain
Frame-based Stereotactic Radiosurgery
Positioning Accuracy
Accuracy and stability of positioning in radiosurgery:
long –term results of the Gamma Knife system.
Heck B et al
Med Phys 2007 Apr; 34(4): 1487-95
Graf Chromic films
densitometric measurements
X: - 0.014+/- 0.09mm
Y: 0.013+/- 0.09mm
Z: - 0.002+/- 0.06mm
MRI-based target definition
X: 0.06+/-0.09mm
Y: 0.04+/-0.09mm
All measured data were within a
sphere of 0.2mm radius
Target delineation: GTV=PTV