Table of Contents Table of Contents
Previous Page  5 / 124 Next Page
Information
Show Menu
Previous Page 5 / 124 Next Page
Page Background

8.

Merieux - US

9.

Abbott - US

10. Aquanal - France

11. Ministry of Agriculture - China

Table 2: Notes on the Eleven Participating Labs

Lab #

Notes

1

Received the protocol with the mistakes, used the manual sample prep (no microwave)

2

Received the protocol with the mistakes and thus used the incorrect lower calibration point.

3

Received the corrected protocol. Did not run prequalification samples first. Results good

4

Received the protocol with the mistakes. They did not run the same calibration curve.

Based on the samples they decided to run 0.1, 0.5, 2.5, 5.0, 10, 25, and 45 mg/L (Protocol

was 0.6, 5, 15, 30, 45, 60, and 75). The highest concentration in their curve was lower than 4

of the 12 samples.

5

Received the protocol with the mistakes, but used the correct lower calibration point.

Passed the practice samples, but both failed when they reran them with the full set of

samples. Paradoxically to that result, this lab had some of the best results in the study

relative to the SLV, especially for the powdered samples which proved to me more difficult

than the RTF samples.

6

Received the protocol with the mistakes, but did not use the incorrect lower calibration

point. They did use their own calibration curve which was 0.5, 1, 2, 5, 8, 10, and 20. The

highest concentration in their curve was lower than 4 of the 12 samples and 2 were right at

the highest concentration standard.

7

Received the protocol with the mistakes, were loaned equipment for the study. Did not run

prequalification samples first.

8

Received the corrected protocol, but still followed the original protocol. Had them

recalculate their results without the lower concentration standard. That had little impact on

the results. Did not run prequalification samples first. One of the two prequalification

samples passed when run with the full set of samples.

9

Received the protocol with the mistakes, though it appears the correct lower calibration

point was used. Despite that the peak area values for that standard appear low. I tried to

contact to recalculate without the lower calibration point, but was unsuccessful with

multiple attempts.

10

Received the protocol with the mistakes, were loaned equipment for the study. Did not use

the incorrect lower concentration standard in their calculations. Passed the practice

samples, but failed (too high) when they reran them with the full set of samples. Overall all

their results were high. Removing this lab from two additional sets of duplicates (data for

this removal was not included reduced one HorRat from 1.97 to 1.64 and other from 2.11 to

1.86.

11

Received the corrected protocol. Did not run prequalification samples first but both samples

passed when run with the full set of samples.

SUMMARY OF MLT

FOR ERP USE ONLY

DO NOT DISTRIBUTE