PDE5-003
Cons/Weaknesses
Low throughput, only applicable to the analytes that can
be ionized at the condition used. In this case, only ESI
positive mode was used.
Supporting Data
General Comments
There are details in this work about the QC, data analysis,
and how to use the method with other techniques such
as NMR to fully characterize other unknowns etc. The
figures and tables are clear. There are some
inconsistencies with the parameters in the method
description and in the appendix, such as mass spectral
acquisition parameters.
Method Optimization
The method optimization is not specifically discussed;
however,the method description indicated that
optimization was performed based on the recommended
solvent selection for the extraction of the analytes and
also the recommend fragmentation parameters etc. of
mass spectrometer.
Performance Characteristics
Analytical Range:
1ug to 5 ug/mL.
LOQ:
Not applicable as a qualitative method
Accuracy/Recovery
Not applicable as a qualitative method
Precision (RSDr)
Not applicable as a qualitative method
Reproducibility (RSDR)
Not applicable as a qualitative method
System Suitability
It is not specifically discussed. Running standards before
running the samples will fulfil part of this requirement.
Safety Review
Is the method safe to use as written?
Yes
Are there any safety concerns with
any component, instrument, reagent
or instruction of the method? If yes,
please clearly state the concerns.
No extra safety concern is identified.
Does the method include all
necessary safety precautions and/or
warning statements? Yes or No. If
No, please suggest potential
precautions or warnings that can be
added to the method for users or
recommend safer option(s) if
yes