Table of Contents Table of Contents
Previous Page  53 / 96 Next Page
Information
Show Menu
Previous Page 53 / 96 Next Page
Page Background

2.

Does the method contain system suitability tests or controls as specified by the SMPR? If no,

please indicate if there is a need for such tests or controls and which ones.

No suitability requirements. A biological control material was used.

3.

Is there information demonstrating that the method system suitability tests and controls as

specified in the SMPR worked appropriately and as expected? If no, please specify.

No, the control was not used for quantitative methods and there were no system suitability

requirements

4.

Based on the supporting information, is the method written clearly and concisely? If no, please

specify the needed revisions.

Yes. It is clear and concise (maybe too concise)

5.

Based on the supporting information, what are pros/strengths of the method?

Good chromatography, a nice range of samples were tested

6.

Based on the supporting information, what are cons/weaknesses of the method?

An internal standard would help. Also, it was not fully validated for both alkaloids

7.

Any general comments about the method?

The method would be greatly improved with the use of a labeled internal standard. Also, the

lack of information about the 7-OH alkaloid needs to be addressed

V.

RECOMMENDATION FOR THE METHOD:

1.

Do you recommend this method be adopted as a First Action and published in the

Official

Methods of Analysis of AOAC INTERNATIONAL?

If no, please specify rationale.

Not at this time, if further information supporting the analytical range, LOQ, LOD, recovery,

RSDr, and RSDR is added for both analytes of interest, then I would consider it (although the

spike recovery is out of range and would likely be improved by the use of labeled internal

standards)