Aloe SLV Method Supplementary Information

Precision: The precision (RSDr) reported in the publication is intermediate precision (different days,

calibration curves, solutions, etc); which is more rigorous in comparison to AOACs guidelines for

repeatability as specified in the SMPR. Threfore, we have complied the repeatability according to AOAC

guidelines below. The repeatability (%RSDr) between replicate samples on the same day for day 1 of the

analysis (4 replicate samples) is summarized below for the samples which contained Aloins above the

quantitation limit.

Repeatability (%RSD)

Aloin A

Aloin B

TM2

3.6

2.8

TM3

0.83

4.3

TM5

1.4

2.1

TM6

0.83

0.74

TM8

0.85

0.51

LOD/LOQ concentration in Samples: The LOD/LOQ reported in the publication is concentration of the

calibration solution. The back calculation for liquid and dry matrices are summarized below:

Liquid samples (as per manuscript):

LOD

aloin A &B: 0.09

µ

g/mL;

LOQ

Aloin A: 0.23

µ

g/mL, Aloin B: 0.21

µ

g/mL.

Powders & capsules:

LOD

Aloin A & B: 0.9

µ

g/g;

LOQ

Aloin A: 2.3

µ

g/g, Aloin B: 2.1

µ

g/g