Background Image
Table of Contents Table of Contents
Previous Page  163 / 620 Next Page
Information
Show Menu
Previous Page 163 / 620 Next Page
Page Background

Page

11

of

14

Table 2-4

Method comparison

New method

997.08

1

999.03

2

Reference 3

Number of sample preparations

and analyses needed to complete

testing

1

3

3

1

Instrument Run time (min)

45

83x3=249

NAP

65

Can test samples with

sucrose:fructan ratio >3:1

Yes

No

Yes

Yes

Requires post-column reagent

addition

No

No

NAP

Yes

Applicable to FOS,

oligofructose, and inulin

Yes

Yes*

Yes*

Yes

Requires extraction step

No

Yes

Yes

Yes

Requires SPE

No

No

No

Yes

Requires dry-down

No

No

No

Yes

Requires knowledge of

commodity type for accurate

results**

Yes

Yes

Yes

Yes

Method bias for any

commodities

No

No

Yes

No

Uses borate trap

Yes

No

No

No

LOQ (% on RTF basis)

0.03%

(90.8-115%

recovery)

0.5%

0.5%

0.014%

(83-101%

recovery)

%RSD (Intermediate

Precision?)

2.04-7.12%

5.79%

1.25%

4.22-8.37%

1

Stated method performance taken from Covance labs website effective 10/15/2010.

2

Stated method performance taken from Covance labs website effective 01/21/2005.

*Method accuracy compromised by usage of set DP factors of 4 and 10 (actual range is from 3.7-30).

**Limits utility at a regulatory level where the type of material may be unknown or testing for off-label

adulteration may be desired. The new method circumvents this via the qualitative ID methodology in the

Appendix. Reference #3 is pursuing an alternate method to circumvent this problem.

Conceptually this methodology is the similar to that of Cuany, et. al. (3) However several time savings are gained from the elimination of

extraction, SPE, and sample dry-down. An additional time savings is generated in a shorter instrument cycle time (45 minutes versus 65

minutes). Some of this is due to the incorporation of a borate trap to the analysis in place of a standard PA1 guard. Borate negatively

impacts HPAEC-PAD chromatography, creating peak distortion and other issues previously observed (3, 5). Furthermore, the addition of

the internal standard (as suggested in 997.08) allows for a non-volumetric dilution and a subsequent improvement in LOQ.

1.

Linearity –

For the 43 standard curves examined no calibration errors greater than + 5% were observed (see figure 2-3). As

expected the largest errors were observed in the lowest level standards. However the pattern of errors does not indicate any

systematic trends.

FOS-04

FOR ERP USE ONLY

DO NOT DISTRIBUTE