equilibrated with 50 mM NaOH for 15 minutes (see Figure 2-1 below). Column and detector compartment are maintained at 20

°

C.

2.

Injection –

A single 4 µL injection of each sample test solution is made. At the start of each sequence, 6 precision injections are

made to ensure that system is stable, followed by the 4 levels of working standard. Samples are bracketed with working

standards after every 13 injections. Autosampler is maintained at 10

°

C.

3.

Electrochemical detector parameters –

This method utilizes the carbohydrate triple waveform referred to in Thermo/Dionex

Tech Note 21, Waveform B (4).

Note this waveform is not appropriate for disposable gold electrodes!

The reference electrode

is set to AgCl mode.

4.

Retention times –

Typically fructose elutes around 10-11 minutes and glucoheptose around 16-18 minutes (see figure 2-2 below).

Figure 2-1

Quantitative Fructan Determination Gradient

Figure 2-2

Working standard chromatogram

Calculations

1.

Fructose Stock Standard –

Concentration is calculated as follows: