Single Lab Validation Report for GOS in Infant Formula and Adult Nutritionals

17

6.

Method:

6.1.

GOS fractionation

Carbopac PA100 column plus guard

Eluent A = 1 M NaOH

Eluent B = 1 M NaAc

Eluent D = Milli-Q

Flow 1 mL/min

Column temperature 20°C

Tray temperature 10°C

Pressure upper limit = 4000 psi

Sample injection 25 µL

Table 5.3: Elution scheme for fractionation

Time (min) Eluent A (%) Eluent B (%) Eluent C (%) Eluent D (%) Curve

0.00

10

4

0

86

5

1.10

10

4

0

86

5

4.00

50

40

0

10

5

8.00

50

40

0

10

5

8.10

10

4

0

86

5

21.0

10

4

0

86

5

6.2.

GOS quantification

Carbopac PA1 column plus guard

Eluent A = 300 mM NaOH

Eluent B = 1 M NaAc

Eluent D = Milli-Q

Flow 1 mL/min

Column temperature 20°C

Tray temperature 10°C

Pressure upper limit = 3000 psi

Sample injection 25 µL

Table 5.4: Elution scheme for quantification

Time (min) Eluent A (%) Eluent B (%) Eluent C (%) Eluent D (%) Curve

0

5

0

0

95

5

20

5

0

0

95

5

33

32

0

0

68

5

35

32

0

0

68

5

39

32

50

0

18

5

43

32

50

0

18

5

45

32

0

0

68

5

48

5

0

0

95

5

63

5

0

0

95

5

GOS-02 SLV

FOR ERP USE ONLY