Parameter

Method Performance

Please

report in units as stated in SMPR!

Weighting

factor for

parameter

Suitability Ranking (1-

3-5)

(select from drop-down

list, 5 = best)

SPIFAN matrices

Applicable to all forms of infant, adult, and/or pediatric

formula (powders, ready-to-feed liquids, and liquid

concentrates).

Any combination of milk, soy, rice, whey, hydrolyzed

protein, starch, and amino acids, with and without intact

protein.

Applicable to all product types.

1

All analytes defined in the

applicability statement are

measured.

Chloride

Chloride is measured.

1

5-500 mg/100g*

minimum from 2.4 mg/100g

(Mehtod can be used in 99.9% salt content with

appropriate dilution factor)

Unknown

Not required

5 mg/100g*

2.4 mg/100g*

Spike recovery (%)

95 - 105 %

10 SPIFAN samples were spiked at 2 levels of

chloride on 3 separate days.

Spike level 1: Average recoveries of 101.6 %

(Recovery Range 99.7- 103.5%)

Spike level 2: Average recovery of 101.9 %

(Recovery Range 100.1 - 104.0 % ).

2

Bias vs SRM NIST 1849a informational value = 710 mg/100g

Value found = 709.5 mg/100g

No bias at 95% confidence level.

3

Average RSDr = 0.87% (10 products)

(RSDr Range 0.43 - 1.34 %; Range evaluated 38.2 -

162 mg/100g RTF)

See table 3 for values

Average RSDr = 1.76% (10 products)

(RSDr Range 0.80 - 3.04 %; Range evaluated 38.2 -

162 mg/100g RTF)

See table 3 for values

b

Units

Bias against established method

Analytical equipment

1

Adequate proof of performance via system suitability

Is there a bias Yes/No ?

Analytical equipment is commonly available in most labs.

3

1

2

a

Concentrations apply to (1) "ready to feed liquids" "as-is"; (2) reconstituted powders (25 g into 200 g water); and (3) liquid concentrates diluted 1:1 by weight.

Did Method Author Consider ERP’s Method Specific

Recommendations

(See web link to specific method

comments):

Decision by ERP

Proprietary equipment

Laboratory safety

Other Considerations

Notes:

Overall Score

Recommendation of ERP

2 years after First Action Status

AOAC SMPR: 2014.015

Limit of detection (LOD)

1

Evaluation of Method Performance vs. SMPR requirements.

Feedback from Users of the Method since being awarded

First Action

Official Methods

Status

Repeatability (RSDr)

Reproducibility (RSDR)

Accuracy/Recovery

Chloride

< 2 %

ADDITIONAL EVALUATION PARAMETERS

Intermediate Reproducibility

(RSDiR)

Chloride

< 4 %

move to Final Action/repeal/remove/expand 2 year term

0

No unique proprietary equipment/accessories are required.

Method does not require any special safety precautions e.g. personal protection from highly toxic solvents.

Limit of quantification (LOQ)

Method Reference #

Method title:

Principle of the method:

Chlorine in Infant Formula and Adult/Pediatric Nutritional Formula by Potentiometric Titration

Sample treats with potassium ferrocyanide and zinc acetate to precipitate protein, acidifies with nitric acid, and titrates using silver nitrate as

titrant. The endpoint of titration determines by the potentiometric method on a silver titrode with silver sulfide coating electrode. Calculate the

chlorine content in the sample via the titrant volume of endpoint.