Background Image
Table of Contents Table of Contents
Previous Page  397 / 620 Next Page
Information
Show Menu
Previous Page 397 / 620 Next Page
Page Background

Parameter

Method Performance

Please

report in units as stated in SMPR!

Weighting

factor for

parameter

Suitability Ranking (1-

3-5)

(select from drop-down

list, 5 = best)

SPIFAN matrices

Applicable to all forms of infant, adult, and/or pediatric

formula (powders, ready-to-feed liquids, and liquid

concentrates).

Any combination of milk, soy, rice, whey, hydrolyzed

protein, starch, and amino acids, with and without intact

protein.

1

All analytes defined in the

applicability statement are

measured.

1

5-200 µg/100g

13.6-1360 µg/100 g

Fluoride

Not provided

1.77 µg/100 g

Fluoride

Not provided

5.86 µg/100 g

Spike recovery (%) (5-25µg/100g): 80–120%; (26-200µg/100g): 90–110%

88.6% (spiked level1), 104.8% (spiked level2)

2

Bias vs SRM

SRM value not provided

SRM value not provided

3

(5-25µg/100g) ≤ 8%; (26-200µg/100g) ≤5%

0.60%

Not provided

(5-25µg/100g) ≤ 15%; (26-200µg/100g) ≤10%

11.50%

b

Units

Bias against established method

Analytical equipment

1

Adequate proof of performance via system suitability

Is there a bias Yes/No ? No certified value provided in SMPR

Analytical equipment is commonly available in most labs.

3

1

2

a

Concentrations apply to (1) "ready to feed liquids" "as-is"; (2) reconstituted powders (1.47 g into 20 g water); and (3) liquid concentrates diluted (5g in 20g water).

Did Method Author Consider ERP’s Method Specific

Recommendations

(See web link to specific method

comments):

Decision by ERP

Proprietary equipment

move to Final Action/repeal/remove/expand 2 year term

0

No unique proprietary equipment/accessories are required.

Method does not require any special safety precautions e.g. personal protection from highly toxic solvents.

Notes:

Overall Score

Recommendation of ERP

2 years after First Action Status

Laboratory safety

Evaluation of Method Performance vs. SMPR requirements.

Feedback from Users of the Method since being awarded

First Action

Official Methods

Status

Repeatability (RSDr)

Reproducibility (RSDR)

Accuracy/Recovery

ADDITIONAL EVALUATION PARAMETERS

AOAC SMPR: 2014.016

Limit of detection (LOD)

1

Intermediate Reproducibility

(RSDiR)

Other Considerations

Limit of quantification (LOQ)

Method Reference #

2014.016

Method title:

Principle of the method:

Single Laboratory Validation for Fluoride Analysis in Infant Formula and Adult Nutritionals: AOAC SMPR 2014.016

Ion Chromatography with suppressed conductivity detection

SMPR Requirement

Analytical Range.

1

SPIFAN ERP Checklist v 1.6

27.06.2013

Fluor-02 - MEF FORM

FOR ERP USE ONLY

DO NOT DISTRIBUTE