Table 2015.06B. Microwave operating parameters: Stages 1 and 2 are

operated sequentially, without removing vessels from the oven

Stage 1 sample digestion

1

Power

100% (1600 W)

2

Ramp to temperature

20min

3

Hold time

20min

4

Temperature

180°C

5

Cool down

20min

Stage 2 sample digestion

1

Power

100% (1600 W)

2

Ramp to temperature

20min

3

Hold time

20min

4

Temperature

200°C

5

Cool down

20min

Total

2 h

Table 2015.06C. Typical ICP/MS parameters for Agilent 7700x

RF power, W

a

1600

RF matching, V

1.8

Sampling depth, mm

9

Extract 1 lens, V

0

Carrier gas, L/min

0.9

Make-up gas, L/min

0.2

Nebulizer (glass concentric)

MicroMist

Spray chamber temperature, °C

2

Interface cones

Ni

He cell gas flow rate, mL/min

4.5

H

2

cell gas flow rate, mL/min

4.2

Nebulizer pump rate, rps

0.1 (0.5mL/min)

Peristaltic pump tubing

White/white, 1.02mm id

Drain tubing

Blue/yellow, 1.52mm id

a

RF = Radio frequency.

Candidates for 2016 Method of the Year