Table 2015.07. Titration Conditions for System Suitability Test (DET U mode) and Sample

Analysis

Module / Parameter

Condition

Start conditions / Pause

15 s

Titration parameters / Measure point density

4

Titration parameters / Minimum increment

10.0 μL

Titration parameters / Dosing Rate

Max. mL/min

Titration parameters / Signal drift

50mV/min

Titration parameters / Equilibrium time

26 s

Titration parameters / Measure input

1

Titration parameters / Stirrer rate

10

Stop conditions / Stop volume

10mL

Stop conditions / Stop measure value

120 mV

Stop conditions / Stop EP

1

Stop conditions / Volume after EP

1mL

Evaluation / EP recognition

greatest

Reports / PC/LIMS

on

Automation / Dripping time

3 s

Automation / Rinsing time

15 s

Automation / Stirring rate

10