Reference methods
Liquid flow measurement is generally used for example for the
measurement of cold potable water, heat metering applications,
gas measurement or fuel dispensers, truck, rail, or ship loading ap-
plications. In addition, industrial processes are sometimes coupled
to flow measurement. For the calibration of all these flow meters,
different reference methods can be used, depending on the required
measurement uncertainty of the meter under test and the application.
Reference meter:
The meter under test is calibrated with another
flowmeter of the same or different construction. Usually more accu-
rate master reference flowmeters which have been specially selected
are used. The metrological behaviour (measurement uncertainty,
repeatability and reproducibility) of those meters is well known and
the meter has a long calibration history. If liquid flowmeters are to
be calibrated, usually calibration rigs with a circulating flow are used.
The medium is taken out of a collection tank and pumped into the
measuring section, where the meter under test is mounted. Within
the defined calibration time the meter under test is compared with the
reference meter. Sometimes the error curve of the reference meter
is corrected during the calibration of the meter under test. Of course
the minimum measurement uncertainty which can be achieved is at
a medium level, because a reference flowmeter must be calibrated
elsewhere with another reference standard with a smaller measure-
ment uncertainty. Depending on the physical principle of the meter
under test, the reference meter method is used for many industrial
calibration processes.
Figure 1
shows a liquid flow calibration rig with
several Electromagnetic reference meters.
Gravimetric weighing system:
With more stringent requirements
of the measurement uncertainty for flowmeter calibration, liquid
flow calibration rigs are used as a reference [6]. The flowing liquid is
measured by a weighing scale (mass), a temperature measurement
(density) and a time measurement see
Figure 2
. The medium is taken
from a collection tank and directly (or with a constant header tank, not
shown in
Figure 2
) pumped into the measuring section. The meter
under test is mounted in the measuring section. The medium passes
through the meter under test and flows through a reference and flow
control valves into the weighing tank. The calibrationmethod shown is
the standing start stopmethod, where the medium is directly pumped
into the weighing tank from the start of the pumps. If the flying start
stopmethod is used, themedium is diverted with a liquid flow diverter
during the measuring time into the weighing tank. In the meantime
the liquid is pumped back into the collection tank [7,8]. The standing
start stop method avoids the influence of the liquid flow diverter,
but of course the error curve influences the result of measurement
uncertainty of the calibration.
Figure 2: Gravimetric weighing system.
Volumetric calibration system: If volumetric measurands have to be
calibrated, then volumetric references are used for the calibration.
The volume of a flowing liquid is compared with a tank or prover
system or any kind of reference volume with a small measurement
uncertainty. As mentioned for the gravimetric weighing systems,
the volumetric calibration systems can be used in the standing start
stop method if a reference volume is used with or without a liquid
flow diverter. For example, a concrete volumetric reference tank is
used in conjunction with level measurement at the local Weights and
Measures authority in Düsseldorf, Germany for the calibration of cold
water meters. Compact systems with a smaller maximum flowrate
than the reference tank systems are realised today as prover systems.
Here the reference volume is defined by a cylinder. A piston displaces
the medium and brings the medium through the meter under test.
These systems are limited for the maximum flowrate and maximum
velocity through the meter under test.
Figure 3: Volumetric reference system.
Validation of measurements
The validation of measurands for liquid flowmeasurement is a more
complex thing compared to the other SI-units. A direct comparison