withanolide in each material varying from 0.15 to 1.10%
w/w (1500–11000 ppm), the acceptable range of recovery is
95–105% (Table 2). Five of the six withanolide analytes were
<95% recovery in at least one of the four materials, but none of
the withanolide analytes was <95% recovery in all four materials.
Recovery of total withanolide content frommaterials spiked
after extraction (WS/06Lot08, WS/05Lot20, RD/1045, and
ERH-46) varied from 91.0 to 99.6% for W-IV, W-V, WF-A,
12-D, and W-A. These results are very similar to the
recoveries when the materials are spiked prior to extraction,
demonstrating that the extraction process is efficient. The
results for W-B were unexpectedly low in this experiment,
at 70.7–72.2% recovery. A review of the experimental
details did not uncover a likely reason for this low recovery,
however it is noted that these values are not typical for this
analyte. If the extraction of W-B were inefficient, one would
expect lower recoveries for the materials spiked before
extraction compared with those spiked after. The observed
lower recoveries of W-B when materials were spiked after
extraction suggests an error in the amount of W-B added, but
this could not be confirmed.
Ruggedness
Table 13 shows the results of ruggedness testing. The
intermediate precision on the sum of withanolides varied from
0.91 to 6.17%. Because the experiment was not performed
as a Youden-type ruggedness trial, the primary source(s) of
error cannot be determined, though the sum of withanolides is
generally higher under condition 2 compared with condition 1.
The differences are small enough, however, that the method is
considered rugged based on intermediate precision.
Stability of Sample Solution
Stability data are presented in Table 14. There was no
significant decrease in withanolide content for any of the
analytes or the sum of withanolides in the sample extract over a
24 h period at 25±2°C.
System Suitability
System suitability parameters were determined throughout
method development and tested during the validation study.
The following suitability parameters define the optimal
performance of the method: (
1
) Injection of mixed standard
preparations in triplicate with every run; (
2
) peak resolution
between withanoside V and withaferin A is >3; (
3
) the tailing
factor for each analyte is <1.5; (
4
) the relative retention times
compared with withanolide A are ~0.70 for withanoside IV,
~0.89 for withanoside V, ~0.92 for withaferin A, ~0.96 for
12-deoxywithastromonolide, and ~1.15 for withanolide B;
(
5
)–the repeatability precision is <2.0% RSD for standards and
samples; (
6
) the retention time window is within 3σ of the mean
retention time of the standard or QC-check sample; and (
7
) the
QC-check sample result is within 10% of the label value. It is
Figure 9. Linearity of response for withanolide B over the full concentration range by (A) linear regression and (B) residual analysis and over
a truncated concentration range by (C) linear regression and (D) residual analysis.
A
B
C
D