withanolide in each material varying from 0.15 to 1.10%

w/w (1500–11000 ppm), the acceptable range of recovery is

95–105% (Table 2). Five of the six withanolide analytes were

<95% recovery in at least one of the four materials, but none of

the withanolide analytes was <95% recovery in all four materials.

Recovery of total withanolide content frommaterials spiked

after extraction (WS/06Lot08, WS/05Lot20, RD/1045, and

ERH-46) varied from 91.0 to 99.6% for W-IV, W-V, WF-A,

12-D, and W-A. These results are very similar to the

recoveries when the materials are spiked prior to extraction,

demonstrating that the extraction process is efficient. The

results for W-B were unexpectedly low in this experiment,

at 70.7–72.2% recovery. A review of the experimental

details did not uncover a likely reason for this low recovery,

however it is noted that these values are not typical for this

analyte. If the extraction of W-B were inefficient, one would

expect lower recoveries for the materials spiked before

extraction compared with those spiked after. The observed

lower recoveries of W-B when materials were spiked after

extraction suggests an error in the amount of W-B added, but

this could not be confirmed.

Ruggedness

Table 13 shows the results of ruggedness testing. The

intermediate precision on the sum of withanolides varied from

0.91 to 6.17%. Because the experiment was not performed

as a Youden-type ruggedness trial, the primary source(s) of

error cannot be determined, though the sum of withanolides is

generally higher under condition 2 compared with condition 1.

The differences are small enough, however, that the method is

considered rugged based on intermediate precision.

Stability of Sample Solution

Stability data are presented in Table 14. There was no

significant decrease in withanolide content for any of the

analytes or the sum of withanolides in the sample extract over a

24 h period at 25±2°C.

System Suitability

System suitability parameters were determined throughout

method development and tested during the validation study.

The following suitability parameters define the optimal

performance of the method: (

1

) Injection of mixed standard

preparations in triplicate with every run; (

2

) peak resolution

between withanoside V and withaferin A is >3; (

3

) the tailing

factor for each analyte is <1.5; (

4

) the relative retention times

compared with withanolide A are ~0.70 for withanoside IV,

~0.89 for withanoside V, ~0.92 for withaferin A, ~0.96 for

12-deoxywithastromonolide, and ~1.15 for withanolide B;

(

5

)–the repeatability precision is <2.0% RSD for standards and

samples; (

6

) the retention time window is within 3σ of the mean

retention time of the standard or QC-check sample; and (

7

) the

QC-check sample result is within 10% of the label value. It is

Figure 9. Linearity of response for withanolide B over the full concentration range by (A) linear regression and (B) residual analysis and over

a truncated concentration range by (C) linear regression and (D) residual analysis.

A

B

C

D