COMMENT PARAMETERS
USER COMMENTS
AOAC OMA
2011.19
Method Name/Manuscript Title
Chromium, Selenium, and Molybdenum in Infant Formula and Adult
Nutritionals by ICP MS
Manuscript Reference
In your experience using the method, does the
method perform according to the method’s
applicability as written?
Yes, the method performs according to the method's applicability.
In your experience with the method, is there
any safety concerns identified while using or
regarding use of the method?
No safety concerns were identified while using the method.
Do you have any information that supports
regarding the reproducibility of the method as
written: If so, please specify and submit
information?
No, the data generated from the lab was submitted to the study
director to evaluate the reproducibility of the method.
Please attach any documentation to support
your feedback, if available.
Based on your experience with the method,
are there any recommended changes to the
AOAC First Action Method as written?
Secondary options for internal standards and analyte isotopes may
greatly improve the versatility of the method towards a variety of
instrument manufacturers and models.
AOAC INTERNATIONAL
Stakeholder Panel for Infant Formula and Adult Nutritionals (SPIFAN)