Steps to initiate radiation

Function of diode D 29

A: Diode working properly

B: Diode disabled (open circuit)

Signal of failure of

the beam

monitoring system

Interlock not activated:

cannot prevent

irradiation

Signal of failure of

the beam

monitoring system

Interlock activated

prevents irradiation

Diode fails (open

circuit) signal not

transferred

Signal is

transferred

through the diode

A

B

Sequence of steps to initiate

irradiation includes a test of

beam monitoring chambers,

but …

… the information about

missing power supply can not

pass through faulty diode …

… interlock is not informed

that monitoring chambers are

missing

… and gives green light to the

next step in the sequence

towards irradiation