AOAC INTERNATIONAL

and subjected to a 3-hour regrowth; Chi-square and POD calculations demonstrate that

500

this difference is not statistically significant. All BAX® System results were confirmed by

501

the culture method, with the exception of one BAX® System positive result for an

502

environmental sample enriched in BPW and subjected to a 3-hour regrowth before

503

testing. Because this result was a low-level positive, and could not be culture confirmed,

504

it is possible that this result is the result of a cross contamination event but it could also

505

be the result of high levels of the co-inoculated contaminating organism (

Citrobacter

506

brakii

) outcompeting the target and preventing culture confirmation.

507

The results for samples tested by the independent laboratory are also summarized in

508

the tables above. For cream cheese samples, analysis by POD and Chi-square

509

demonstrate that the difference in results between the two methods (14/20 positive for

510

the BAX® System method and 10/20 for the reference method) was not statistically

511

significant. For environmental samples, the BAX® System returned results identical to

512

the reference culture method.

513

514

515

17. LOT-TO-LOT AND STABILITY STUDY

516

517

Methodology

– BAX® System Real-Time

Salmonella

test kits from three lots with different

518

expiration dates were evaluated in experiments to establish lot-to-lot consistency and

519

determine a reasonable shelf-life using both accelerated and non-accelerated storage

520

conditions (see Table 4 below). One

Salmonella

strain was assayed ten times per time

521

point per test kit using cells diluted to one order of magnitude over the product’s claimed

522

sensitivity level (i.e. 10

5

cfu/mL) by the BAX® System assay. Additionally, one non-target

523

strain was tested six times per strain per time point diluted to 10

8

cfu/mL. Also, for each

524

condition, for each time point, two non-spiked lysis buffer controls were tested.

525

Based on these results and applying the Q

10

rule of the Arrhenius equation, a four year

526

shelf life would be supported.

527

Predicted Stability = Accelerated Stability X 2

t/10

528

529

In this case: Stability of a product at 37°C is > 147 days.

530

Recommended storage temperature is 4°C and n = (37 - 4)/10 = 3.3

531

Dupont BAX Salmonella PTM Report

PTM Certification No. 081201 (08/07/2012)

For Expert Review Panel Use Only

Do Not Distribute