OMA-2013-Apr-XXX

compared to traditional immunoassays. The SPR serves as the solid phase as well

1

as the pipetting device. The interior of the SPR is coated with proteins specific

2

for

Listeria

receptors adsorbed on its surface. Reagents for the assay are ready-to-

3

use and pre-dispensed in the sealed reagent strips.

4

5

All of the assay steps are performed automatically by the instrument. The

6

reaction medium is cycled in and out of the SPR several times. Part of the

7

enrichment broth is dispensed into the reagent strip. The receptors present will

8

bind to the proteins specific for

Listeria

receptors, which are coated on the interior

9

of the SPR. Unbound sample components are washed away. Protein conjugated

10

with alkaline phosphatase are cycled in and out of the SPR and will bind to any

11

Listeria

receptors which are in turn bound to the specific protein on the SPR wall.

12

.

13

Further wash steps remove unbound conjugate. During the final detection step,

14

the substrate (4-Methylumbelliferyl phosphate) is cycled in and out of the SPR.

15

The conjugate enzyme catalyzes the hydrolysis of this substrate into a fluorescent

16

product (4-Methylumbelliferone), the fluorescence of which is measured at 450

17

nm.

18

19

At the end of the assay, the results are analyzed automatically by the instrument

20

which generates a test value for each sample. This value is compared to a set of

21

stored standards (thresholds) and each result is interpreted (positive, negative).

22

23

1.3

Summary of the LMX PTM/Precollaborative Study

24

25

The VIDAS LMX method was validated according to AOAC Guidelines (2002)

26

in a harmonized PTM study. The aim of this study was to demonstrate that the

27

VIDAS LMX method could detect

L. monocytogenes

in a variety of foods as

28

claimed by the manufacturer.

29

30

In the PTM study, there were 7 matrices tested using a 26-hr proprietary

31

enrichment broth protocol for the VIDAS LMX method and the appropriate

32

reference method: deli-ham (USDA/FSIS ML

G 1 )

, processed cheese (AOA

C 2 )

,

33

vanilla ice cream (AOAC), cooked shrimp (FDA-BA

M 3 )

, smoked white fish

34

(FDA-BAM), frozen spinach (FDA-BAM) and peanut butter (FDA-BAM).

35

Additionally, there were five ready-to-eat matrices (hot dogs, deli turkey, deli

36

ham, fermented sausage, and liver paté) evaluated using the VIDAS LMX

37

method, the Health Canada MFHPB-3

0 4

method and the USDA/FSIS MLG

38

method.

39

40

1

USDA/FSIS Microbiological Laboratory Guidebook, 8.07, August 2, 2009 effective date

2

AOAC 993.12

3

FDA Bacteriological Analytical Manual, Chapter 10, January 2003 version

4

Health Canada MFHPB-30,

http://www.hc-sc.gc.ca/fn-an/res-rech/analy-meth/microbio/volume2/mfhpb30-

eng.php

DRAFT DOCUMENT

4