Chem, Res. Center Crete, IL – Validation of a LC-MS/MS Method for Vitamin K Analysis

Silliker Laboratories, Chemistry Research Center, Date: 1/30/15

21

RSD of the replicates in the range of 1.7-13.0 %. The lower relative level of the analyte in samples may

be associated with the observed higher variance in the analysis of the replicates.

Trans –vitamin K1 was found to be in the range of 2.1 – 10.3 µg/100g in the analyzed samples. % RSD

of the trans-vitamin K1 analysis was estimated to be under 8% in replicates of many of the analyzed

samples. But % RSD in the trans-vitamin K1analysis of the replicates of some of the samples gave

higher % RSD values in their replicate analysis. . The sample with concentration of Vit K1 higher than

10 µg/100g provided % RSD value of 5% as stipulated in the SMPR.

A summary of the Results for the averages of repeatability/interday-precision (%RSD

r

) for each of the

sample are presented Table 6.

Table 6: Analysis of different SPIFAN samples in replicates on multiple

days – Results of all replicates including those analyzed on different days in replicates

Average of Analysis of replicates on All Three Days

Sample ID

Paramtrs

Trans

Cis

Total

(trans +

cis)

mcg/100g of RF sample as is or a powder

sample reconstituted in to a liquid

SPIFAN

Sample #6

Mean

10.1

1.4

11.5

Stdev

0.6

0.1

0.7

% RSD

6.2

10.4

6.3

n

8

8

8

SPIFAN

Sample #7

Mean

2.2

1.1

3.3

Stdev

0.3

0.1

0.4

% RSD

15.3

10.8

13.0

n

7

7

7

SPIFAN

Sample #8

Mean

5.9

1.1

7.0

Stdev

0.7

0.1

0.8

% RSD

11.8

9.3

11.2

n

8

8

8

SPIFAN

Sample #9

Mean

7.1

1.0

8.1

Stdev

0.4

0.1

0.5

% RSD

5.8

10.1

6.1

n

9

9

9

SPIFAN

Sample #10

Mean

2.4

0.3

2.6

Stdev

0.3

0.0

0.3

% RSD

12.8

11.1

11.0

n

9

9

9

SPIFAN

Sample #11

Mean

2.2

0.4

2.6

Stdev

0.2

0.0

0.2

% RSD

7.4

10.9

6.3

VitK-01 w/SLV