Report: SPIFAN FOS-01 Single Lab Validation (SLV) for fructan in Infant Formula
and Adult Nutritionals.
1. Introduction.
During the September session (2014) at Boca Raton, the SPIFAN GOS/FOS working
group approved the AOAC 997.08 method for further evaluation for a SLV for the FOS
determination.
The AOAC 997.08 method is an enzymatic, chromatographic method for the
determination of fructans in food and feed products. At the time of approval of the
AOAC method (1997) a collaborative study was performed on samples with different
matrices. This method is applied on a daily basis in our lab. The method gives freedom
to apply a chromatographic technique as long as the determination of the sugars can
be done accurate and without interference. Depending on the matrix of the sample, we
decide which chromatographic technique is performed, being GC, HPLC or HPAEC.
For dairy samples our preferred method is GC because of the complexity of the matrix.
A yearly proficiency test is organized by DLA in Germany where different labs are
asked to analyze for the fructan content of a double blind commercial sample. More
info can be found on
http://www.dla-lvu.de.The matrix changes every year. Our lab
participates from 2008 on.
2. Methodology
2.1Samples for SLV.
A Test Materials Kit, ordered at AOAC, and Covance provided different products of
infant and adult nutritionals like listed in the table below. As a first step the 3 placebo
samples and the 12 fortified samples were screened in duplicate for the presence of
fructans.
3 fortified samples (J4G-00019, J4G-00974 and J4G-00980) and also one of the
placebo samples J4G-00013 were found to contain fructans.
The pure products Orafti HP(inulin), Orafti P95(FOS) and Nutraflora P95(FOS) were
screened in duplicate by 3 different chromatographic techniques, being GC, HPAEC
and HPLC. Influence of the applied chromatographic technique is negligible as all
results for the fructan content are on the same level.
The setup of the recovery (spiking) scheme was done in such a way that the 3 different
raw products were used over the whole range.