rapid strand (containing 4/5/6 seeds) inserted into well-

chamber. They are: total activity 2,20 mCi

strand of 4 seeds

produce 121,4 pC/10s corrected for pressure and

temperature reading of electrometer, overall activity 3,28

mCi

strand of 6 seeds

gives 181,6 pC/10s corrected for pressure

and temperature reading of electrometer.

Conclusion

Described quality control procedure provide confidence

that total seeds radioactivity of implanted into patient in

LDR brachytherapy procedure correspond to intended

value. Sterile conditions for rapid strands are guaranteed

and therefore procedure can be accepted for intra-op

using. The exposure dose received by personal during this

procedure is negligibly small.

EP-1795 A novel MRI markers system in applicator

reconstruction for brachytherapy

A. Otal

1

, J. Richart

1

, C. Domingo

1

, S. Rodriguez

1

, P.

Rodriguez-Meijide

2

, I. Fotina

2

, M. Santos

1

, J. Perez-

Calatayud

3

1

Hospital Clinica Benidorm, Radiotherapy, Benidorm,

Spain

2

Eckert & Ziegler BEBIG GmbH, Marketing, Berlin,

Germany

3

Hospital Clinica Benidorm/Hospital La Fe,

Radiotherapy, Benidorm/Valencia, Spain

Purpose or Objective

ABS and GEC-ESTRO recommend magnetic resonance

imaging (MRI) in T2-sequence as preferable modality in

image-guided

brachytherapy

for

gynecological

malignancies. On the other hand, widespread availability

of the applicator libraries in the brachytherapy treatment

planning systems (TPS) allows global approach to the issue

of the applicator modeling. The aim of this work is to

present a method to incorporate complete applicator

geometries in SagiPlan, (not only single channels), based

on the technical data available from manufacturer and

then develop a marking system which allows recognizing a

defined applicator (Vienna-type applicator of Eckert &

Ziegler BEBIG) in the MR image series with utmost possible

precision for the library reconstruction.

Material and Methods

The investigated applicator model is Vienna-type

applicator, consisting of tandem with a ring for possible

interstitial component from Eckert&Ziegler BEBIG (Berlin,

Germany) together with a model available in the library of

the SagiPlan® TPS from Eckert&Ziegler BEBIG.

3D model of the applicator was reconstructed

independently, using open-source modeling software

FreeCAD

(www.freecad.org

).

The presented reconstruction technique is based on the

placement of three vitamin A capsules, clearly visualized

in T2 MRI, embedded in the material layer giving enough

contrast for the capsule recognition. The position of the

marking capsules with respect to the applicator geometry

is known precisely, due to placement of the attachments

in 3D model reconstructed in FreeCAD and verified against

technical specifications of the applicator. The accessory

attachments to the applicator are produced on the 3D

printer from polylactide polymer material (PLA) and carry

the marking capsules inside.

Results

In order to validate the accuracy of the reconstruction

with use of the model described above, we acquired T2

MRI sequence of the Vienna applicator in water with

accessory attachments in place. The images were entered

in the SagiPlan® TPS in order to perform the applicator

reconstruction based on the markers and pre-defined

template, based on the 3D applicator model, loaded in

SagiPlan®.

The presented technique could be expanded to any rigid

applicator and other image modalities, under condition of

the adequate selection of the markers. It is necessary to

mention, that positioning of the accessory attachments on

the applicator is very easy and does not affect integrity or

other characteristics of the product.