Table of Contents Table of Contents
Previous Page  537 / 1096 Next Page
Information
Show Menu
Previous Page 537 / 1096 Next Page
Page Background

S522

ESTRO 36

_______________________________________________________________________________________________

Material and Methods

A decommissioned rectal retractor was modified by

drilling a small hole to allow a microMOSFET to be

inserted. The MOSFET was commissioned measuring

energy dependence and angular dependence of response

for the range of source-MOSFET positions expected in

cervix brachytherapy treatments. Standard and conformal

cervix plans covering the range of applicator sizes and

geometries used in clinical treatments were delivered in a

water phantom. The MOSFET was monitored during

treatment delivery and measured doses compared to

treatment planning system (TPS) calculated doses for the

total plan and for ring and inter-uterine tube (IUT)

individually.

Results

Corrections were applied for energy dependence response

(6% variation between 1 and 8 cm source-MOSFET

positions) and angular dependence of response (up to 8%

under response for the largest polar angle of 170°). Total

plan measurements agreed with TPS calculated doses

within 3.1% - 7.8% for 30° and 60° applicators but

measured 16% -24% high for 45° applicators (k=2

uncertainty was estimated as 14% for total plan

measurements). Separate analysis of ring and IUT

measurements similarly showed good agreement for all

cases except the 45° IUT for which measurements were on

average 55.3% higher than expected. For the 45° IUT the

MOSFET position is directly in line with the source cable

and longitudinal source axis based on the source positions

assumed by the TPS (see figure). A combination of a small

rotation of the source relative to the IUT axis and

deviation of the actual source position from the centre of

the IUT could explain the measurement difference. To

verify this, treatments for the 45° applicator were re-

measured with the MOSFET taped to the outside of the

rectal retractor in a position that was not aligned to the

IUT and measured doses agreed within 8%.

Conclusion

In vivo dosimetry for cervix brachytherapy would be

feasible if commercial rectal retractors were designed to

allow a dosimeter to be inserted. However it is important

to avoid dosimeter positions aligned with the source

longitudinal axis as this is a region of high dose

uncertainty.

PO-0943 Evaluation of a recent in vivo dosimetry

methodology for HDR prostate BT using MOSFET

detectors

R. Fabregat Borrás

1

, S. Ruiz-Arrebola

1

, E. Rodriguez

Serafín

1

, M. Fernández Montes

1

, A. García Blanco

2

, J.

Cardenal Carro

2

, J.T. Anchuelo Latorre

2

, M. Ferri Molina

2

,

A. Kannemann

2

, D. Guirado

3

, P.J. Prada

2

1

Hospital Universitario Marqués de Valdecilla,

Radiophysics, Santander, Spain

2

Hospital Universitario Marqués de Valdecilla, Radiation

Oncology, Santander, Spain

3

Hospital Universitario San Cecilio, Radiophysics,

Granada, Spain

Purpose or Objective

In vivo dosimetry (IVD) applied to HDR BT treatments

allows to monitor real dose delivered to clinically relevant

areas. MOSFET detectors are the most suitable devices for

this task because of their tiny dimensions, which enables

their introduction into identical needles to those used in

treatments. However, these type of detectors show

responses depending on source-to-detector angle and

distance. Mathematical models describing these

dependences can be obtained from a correct detector

characterization. Applying these models on the

measurements should minimize the impact of those

dependences, improving precision and accuracy. The

purpose of this study was to evaluate clinical data of IVD

applied to HDR prostate BT using MOSFET TN-502RDM from

Best Medical Canada with the Ir-192 Vr2 source contained

in Flexitron aferloader (Nucletron-Elekta) and

mathematical models describing those dependences

obtained in a previous characterization work.

Material and Methods

Clinical data were taken from five patients suffering from

prostate cancer. One to three measuring points were

taken for each patient, where the MOSFET were

positioned. Anatomical areas measured were

neurovascular bundles, bulbourethral area and

periurethral area. Nine measuring points were taken and

evaluated.

Real time ultrasound image guided technique was used to

implant the treatment needles. An additional needle was

needed for each measuring point. Oncentra Prostate

4.2.2.4. was used to calculate the treatment planning

following a standard procedure. Subsequently,

coordinates of measuring points and dwell positions were

taken as well as dose contribution of each dwell position

to each measuring point.

After irradiation, mathematical models were applied on

measured dose. Table 1 shows the three models

considered, their parameters and the goodness of fit. TPS

dose, direct MOSFET measured dose and calculated dose

after applying the mathematical models on direct MOSFET

dose were evaluated.

Results

Figure 1 shows the results normalized to TPS d ose. All

measurements suffer an approach to TP S dose after

applying the mathematical models. The ave rage value of

percentage difference between TPS dose and direct

measured dose was 15% while the average percentage

difference after applying the mathematical models

decreases to 9% without any point exceeding 20%.

Estimated global uncertainty associated to these

corrected measurements were into the range 3.7-4.3%.