CT scan used for radiotherapy treatment planning can

contain temporary gas pockets inside the target volume

for patients with tumours in the pelvis that later disappear

during the course of radiotherapy. In this study we are

interested to explore the dosimetric impact of applying

the type C dose calculation algorithm for patients treated

in the pelvic area using VMAT, where gas pockets appear

and disappear in the rectum.

Material and Methods

Ten clinical cervix cancer patients were selected for this

study. The patients had different sizes of gas pockets on

the planning CT. The treatment plans were optimized and

calculated using one type B and one type C dose

calculation algorithm (Eclipse, AAA and Acuros XB(Dose to

Medium), respectively). Gas pockets of these patients

were delineated on the planning CT and the contours and

CT data was duplicated. On the first series the original CT

the HU were kept as is, and on the second CT set the HU

on the gas pockets were overridden with HU zero. Thus,

we simulate a worst-case scenario where the gas pocket is

present on the planning CT but disappear during the

course of treatment. The original treatment plan

optimized using the type C algorithm was recalculated on

both CT data sets. The volume and maximum diameter of

these gas pockets were measured and the difference in

average and maximum dose in these pockets were

calculated.

Results

The average volume of the pockets was 28.6 cm

3

[range:

1.7-77.7 cm

3

] and the average maximum extent of the gas

pocket was 4.2 cm [range: 2.8–5.3 cm]. Volumes up to 20

cm

3

show a decrease about 1 % in the average and

maximum dose to the delineated air pocket for type C and

an increase of less than 1% for type B algorithm. For

volumes larger than 20 cm

3

the average and max dose to

the delineated air pocket increases with more than 5% and

up to 30%, respectively for the type C algorithm. The type

B algorithm shows a decrease up to 2% in average dose and

a small increase in maximum dose.

Conclusion

Gas pockets above a volume of 20 cm

3

in the initial CT scan

can induce dose hotspots during the actual treatment,

when using the type C dose calculation algorithm for

treatment planning and when deriving dose to medium for

VMAT plans. This could increase risk for radiation toxicity

in the worst case scenario. This effect is smaller, opposite

and probably clinically negligible for the type B algorithm.

EP-1463 MONET: an accurate model for the

evaluation of the ion dose in water

A. Embriaco

1,2

, E.V. Bellinzona

1,2,3

, A. Fontana

2

, R.

Alberto

1,2

1

Pavia University, Physics, Pavia, Italy

2

Istituto Nazionale di Fisica Nucleare INFN, Physics

Departiment, Pavia, Italy

3

Ludwig Maxmilians University, Medical Physics, Munich,

Germany

Purpose or Objective

MONET (Model of ioN dosE for Therapy) is a code for the

computation of the 3D dose distribution for protons in

water. MONET accounts for all the physical interactions

and is based on well known theories.

Material and Methods

The first part is the evaluation of the lateral profile.

Our model is based on the Molière theory of multiple

Coulomb scattering. To take into account also the nuclear

interactions, we add to Molière distribution the Cauchy-

Lorentz function, where two free parameters are obtained

by a fit to a FLUKA simulation (Bellinzona et al., PMB

2016).

The next step is the passage from the projected lateral

distribution to a 2D distribution. The projected

distributions are uncorrelated but not independent and we

have to use the Papoulis theorem that allows, in case of

cylindrical symmetry, to rebuild the radial distribution

starting from projected one. We have implemented the

Papoulis algorithm in the code.

The second part is the study of the energy deposition in

the longitudinal profile.

We have implemented a new calculation of the average

energy loss that is in agreement with simulations and other

formulas published in the literature.

The inclusion of the straggling is based on the convolution

of energy loss with a Gaussian function.

In order to complete the longitudinal dose profile, also the

nuclear contributions are included in the calculation using

a linear parametrization with only two free parameters for

energy.

The total dose profile is calculated in a 3D mesh by

evaluating at each depth the 2D lateral distributions and

by scaling them at the value of the energy deposition.

Results

We have compared MONET results with the FLUKA

simulations and we have obtained a good agreement for

different energy of protons in water.

We have reproduced a lateral scan as a sum of many pencil

beams in order to estimate the accuracy of the model

focusing on the tails of the distribution that give rise to

the low-dose envelope. Also in this case, the agreement

between MONET and FLUKA is good.

We have also estimated the calculation time: for each

depth, it is about 2 seconds for the single beam and 4

seconds for the beam scan.

Conclusion

The advantages of MONET are the physical foundation, the

fast calculation time and the accuracy.

A possible development of this study is the creation of a

dose database of clinical interest and an online fast dose

evaluation tool. In the next future, we would like to

extend MONET to the case of Helium beam and other ions.

Preliminary results for helium ion will be shown.

EP-1464 Investigation on beam width tolerances for

proton pencil beam scanning

B. Ackermann

1

, S. Brons

1

, M. Ellerbrock

1

, O. Jäkel

1,2

1

Heidelberg Ion Beam Therapy Center HIT, Medical

Physics, Heidelberg, Germany

2

German Cancer Research Center, Medical Physics in

Radiation Oncology, Heidelberg, Germany

Purpose or Objective

Beside beam spot position and proton range, the beam

spot width is one of the central parameters for the correct

application of a proton therapy plan utilizing pencil beam