Table of Contents Table of Contents
Previous Page  890 / 1096 Next Page
Information
Show Menu
Previous Page 890 / 1096 Next Page
Page Background

S874

ESTRO 36

_______________________________________________________________________________________________

respectively. DIFGI is benchmarked against Varian’s RPM

until final validation of the device, but it is compatible

with all treatment units and CTs.

Fifteen left-sided breast cancer patients have been

recruited until now. If heart constraints can’t be fulfilled

in free-breathing (FB), then patients are trained and

undergo a second CT scan in DIBH using the DIFGI.

The stability, repeatability, reproducibility and reliability

of the method are studied. Two radiopaque markers, one

on the mediastinum tattoo and another along the back,

serve as a reference to measure breath amplitude (Fig.2).

The stability and repeatability are measured on the DIBH

CT scan. The reproducibility mean value, systematic, and

random errors are determined by using daily kV images

and weekly CBCTs. The reliability of the device is

calculated as the failure ratio compared to RPM.

We also analyse Dmean, V30 (cm

3

), and V25 (%) for the

heart in both techniques.

Results

Stability and repeatibility are below 1.7 and 3.3 mm in all

cases, respectively. Repoducibility mean value is 1.7 mm,

systematic error is 0.5 mm, and random error is 0.9 mm.

DIFGI reliability is 95%. All failures are human errors

occurred during the learning period.

Dosimetric benefits compared to FB for the heart are: 3.0

vs 6.7 Gy for mean dose, 14.9 vs 53.4 cm

3

for V30, and 2.8

vs 9.5% for V25.

Conclusion

DIFGI is a simple, friendly, low-priced external

respiration-monitoring device compatible with all

treatment units and CTs. The preliminary results of the

stability, repeatability, reproducibility, and dosimetrical

benefits are encouraging. The reliability of the device

depends on human intervention so we plan to interlock it

with the treatment unit.

EP-1617 Reproducibility and stability of vmDIBHs

during breast cancer treatment measured using a 3D

camera

M. Kusters

1

, F. Dankers

1

, R. Monshouwer

1

1

Radboud university medical center, Academic

Department of Radiation Oncology, Nijmegen, The

Netherlands

Purpose or Objective

To accurately perform voluntary moderately deep

inspiration breath hold (vmDIBH) radiation therapy it is

essential to determine the position of the chest wall at the

start of treatment and to monitor deviations during

treatment.

An in-house developed real-time automated monitoring

system of the respiratory motion is implemented to verify

the reproducibility and stability of the vmDIBH during

breast cancer treatments.

Material and Methods

Patients with left sided breast cancer are guided to

perform vmDIBH assisted by verbal instructions and an

additional aid called the ‘breathing stick’ [1].

A 3D Kinect v2 camera (Microsoft, USA) was mounted in

the treatment room to visualize the patient on the

treatment couch. Software was developed to track and

visualize the anterior–posterior motion of a small area of

the surface of the thoracic wall in real time, allowing RTTs

in the treatment room to verify the reproducibility and

stability of the breath holds during treatment.

The data of ten patients was analysed for reproducibility

and stability. The formulas for reproducibility and stability

were derived from Cerviño et al. with minor adaptations

[2]. For reproducibility the standard deviation of the mean

of each DIBH level was used. For stability all breath holds

were fitted by first order polynomials, the slopes were

multiplied by their breath hold lengths to find a range and

all these ranges were averaged.

Results

A typical example of reproducible and stable vmDIBHs

during a treatment fraction of a patient is shown in Figure

1.

Figure 1:

Typical example of five reproducible and stable

vmDIBHs for one treatment fraction

The analysed reproducibility and stability of vmDIBH

treatment for then patients using the breathing stick is

shown in Figure 2. The median reproducibility and stability

were 0.9 mm and 1.1 mm, respectively.

Figure 2:

Clinical measurements of the reproducibility and

stability of the vmDIBHs for ten patients

Conclusion

The reproducibility and stability of the chest wall can be

accurately measured using the in- house developed

monitoring system. vmDIBH in combination with the

breathing stick shows good stability and reproducibility

which are comparable to the results in the study of

Cerviño et al. [2].

In this work the current results are limited to ten patients;

we continue to, acquire more data for future analyses.

The breathing stick is routinely used in our clinic;

currently we use the breath hold monitoring system to test

whether using this tool has an added value.