Input data – “general” to PB electron

calculations

Measured Depth–Dose Distributions

o

Depth dose distributions for electron beams of a given E

0

are extremely

dependent on the particular beam transport system (scattering foils or

scanning system, applicator design. etc.) of the user’s accelerator. The

Fermi–Eyges model requires central-axis depth– dose distributions for

a number of different rectangular field sizes –

g

(

z

)

Dose Profiles

o

A dose profile is required in order to derive the weighting factor W (x,y).

This should be measured at or near d

max

in a principal plane across the

larger field dimension for each field for which the CAXD (z ) is required.

Mean Energy at the Surface, E

0

The Initial Angular Spread

o

This parameter can be determined from in-air measurements of the

penumbra at different distances from the collimator using film

(Hogstrom et al. 1981).