Fig. 4.

NEBNext Ultra DNA reagent calculator.

Fig. 5.

Automated library construction plate layout.

Fig. 6.

E. coli

K12 gDNA control libraries created with the NEBNext

Ultra DNA automated method.

In addition to the user interface, the automated NEBNext

Ultra DNA method provides the user with an HTML-

driven reagent calculator that provides the user with the

final volumes of all of the reagents and master mixes

required on the deck, as well as instructions on how to

generate the various master mixes—based upon the

number of samples to be processed and the amount of

the workflow that the user wishes to pursue. An image

of the reagent calculator is presented in Figure 4.

Results

Sixteen genomic DNA (gDNA) samples from

A. thaliana

and

H. sapiens

, 2

C. elegans

amplicon pools, 2

H. sapiens

ChIP DNA, and 4 bacterial gDNA samples were supplied

by various laboratories at Indiana University, Bloomington,

for Biomek automated library construction at Indiana

University. In addition,

E. coli

K12 gDNA was supplied

by New England Biolabs as a positive control. The 200 ng

aliquots for each of the gDNA samples were sheared

to an average size of 400 bp (data not shown) and arrayed

in the sample plate as shown in Figure 5.

Libraries were constructed using the NEBNext Ultra DNA

automation method utilizing the 400–500 bp insert size

selection option. Eight cycles of PCR enrichment were

used to amplify the libraries using an off-deck thermocycler.

Following analysis on the Agilent 2200 TapeStation, all

of the libraries processed were then sequenced on an

Illumina MiSeq

®

using a 2x300 cycle paired end run.

For the purposes of this document, we concentrated

our analysis on the six

E. coli

K

1

2 control libraries, which were

assayed on the Agilent 2200 TapeStation using D

1

000

ScreenTape (PN# 5067-5582). The electropherograms

for all 6

E. coli

K

1

2 control libraries are presented in Figure 6.

Data analysis was performed at New England Biolabs

using a local instance of Galaxy. For each of the 6

E. coli

K

1

2 control libraries, over

1

.4 million pass filter reads

for each library were generated by the MiSeq run. Pass

filter read counts are presented in Figure 7. Reads were

then trimmed using SeqPrep prior to mapping back to

the

E. coli

K

1

2 MG

1

655 reference genome using Bowtie

(version 2.

1

.

1

0). Less than

1

% of the reads from the

E.

coli

K

1

2 control libraries failed to map back to the

E. coli

K

1

2 MG

1

655 reference genome. We also observed low

percentages of chimeric reads, PCR duplicates, and

mismatched reads. These metrics indicate that the libraries

were high-quality libraries. Data on basic library quality

metrics are presented in Figure 8. Additionally, we

investigated the average size of the library inserts of

the

E. coli

K

1

2 control libraries utilizing the library

mapping data. As shown in Figure 9, the average size

inser ts of the

E. coli

K

1

2 control libraries was

approximately 400 bp.

AAG-350TCH08.14-A