Table of Contents Table of Contents
Previous Page  110 / 453 Next Page
Information
Show Menu
Previous Page 110 / 453 Next Page
Page Background

3M MDA2 E. coli O157:H7 Collaborative Study

May 2016

OMA-2016-May-XXXX

DRAFT DOCUMENT

6

e)

Most Probable Number (MPN) Analysis.

The level of

E. coli

O157 will be determined by MPN

1

on the day of analysis by evaluating 5 x 650 g, 20 x 325 g (using the reference method test

2

portions) or 5 x 325 g and 5 x 160 g inoculated test samples. Each test portion will be analyzed

3

by the reference method procedure. The number of positives from the 3 test levels was used to

4

calculate the MPN using the LCF MPN calculator available at the AOAC RI website.

5

6

3)

3M™ Molecular Detection Assay 2 -

E. coli

O157 Method

7

8

a)

Applicability

9

10

For the detection of

E. coli

O157

from: raw ground beef (325 g), frozen blueberries (25 g),

11

sprouts (25 g), fresh baby spinach (200 g).

12

13

b)

Test Kit Information

14

a)

Kit Name.—

3M™ Molecular Detection Assay 2 –

E. coli

O157 (including H7)

15

b)

Catalog Number

.—

MDA2ECO96

16

17

c)

Test Kit Reagents -

3M MDA2 -

E. coli O157

(including H7) (96 tests)

18

a)

Lysis Solution (LS) tubes

. – 96 (12 strips of 8 tubes)

19

b)

E. coli O157 Reagent tubes

. – 96 (12 strips of 8 tubes)

20

c)

Extra caps.

– 96 (12 strips of 8 caps)

21

d)

Reagent Control (RC).

– 16 (2 pouches of 8)

22

e)

Quick Start Guide

23

24

d)

Additional Supplies and Reagents

25

a)

3M Molecular Detection System (instrument), power supply, cord, USB cable and software

26

b)

3M Molecular Detection Speed Loader Tray

27

c)

3M Molecular Detection Chill Block Tray and Chill block insert

28

d)

3M Molecular Detection Heat Block Insert

29

e)

3M Molecular Detection Cap/Decap Tool [Reagent]

30

f)

3M Molecular Detection Cap/Decap Tool [Lysis]

31

g)

Empty lysis tube rack

32

h)

Empty reagent tube rack

33

i)

3M™ Enrichment Pouch with 225mL of Buffered Peptone Water (ISO formulation) and filter,

34

available from 3M #QEBPWI225F

35

j)

Matrix Control, available from 3M #MDMC96NA

36

37

e)

Additional Media for Confirmation

38

a)

Modified Rainbow agar (mRBA)

39

b)

Tryptic soy agar with 5% sheep blood (SBA)

40

c)

Dynal antibody coated paramagnetic beads

41

d)

Biochemical confirmation supplies

42

e)

Serological confirmation supplies

43

44

f)

Apparatus

45

a)

Pipettes

. – Capable of 20µL

46

b)

Multi-channel pipette

. – Capable of 20µL

47

c)

Sterile pipette tips.

– Capable of 20µL

48

OMAMAN-35 B: Collaborative Study Protocol

ERP USE ONLY

January 2017

AOAC Research Institute

Expert Review Panel Use Only