Table of Contents Table of Contents
Previous Page  528 / 696 Next Page
Information
Show Menu
Previous Page 528 / 696 Next Page
Page Background

Silliker

Chem, Res. Center Crete, IL – Report of A Validation of LC-MS/MS Method for Folate Analys

24

The above results of total folate as folic acid (sum of the detected folate compounds after calculation as

folic acid) in the SPIFAN samples obtained using the trienzyme extraction including the conjugase

treatment for all 12 SPIFAN samples and four check samples are based on analysis on 3 separate days

in triplicates (SPIFAN samples). Averages and interday-repeatability value/precision (%RSD) for the

analysis are presented.

Total folate values obtained for the reference materials are within the specified range of total folate for

respective sample.

Total folate concentration determined in the reconstituted liquid samples is in the range of 16-34

mcg/100g and in the range of 16.5 to 68 mcg/100g in the RTF samples. The inter-day-precision or

repeatability (%RSD) obtained from the analysis in of all the analyzed SPIFAN II samples was found

to be less than 5%. The precision RSD

r

requirement of the folate SPIFAN SMPR (2011.006) is 11% or

less for concentration range 0.5 to 21.4 mcg/100g and 7% or less for higher concentration samples

(21.5-85 mcg/100g). Thus the method meets the precision requirement of the Folate SMPR for all the

SPIFAN samples (analyzed on 3 days in triplicates).

Intermediate precision or inter-day precision as % RSD not clarified in the SMPR requirements. Inter-

day precision as % RSD obtained values are well below the reproducibility requirement (as %RSD

R

) of

the SMPR (< 16%). The range of the inter-day precision (as % RSD) obtained by the method are 7 –

12.7% .

Averages and Precision of folate analysis in SPIFAN II samples using protease and amylase but

no conjugase treatment:

Nine beverages prepared from the respective powdered SPIFAN samples and three RTF samples were

analyzed for all the targeted folate compounds. The folate in these samples was extracted using

protease and amylase but the conjugase treatment was excluded during the extraction. Results in the

reconstituted samples were calculated and reported on reconstituted beverage basis (mcg/100g). The

folate compounds in the RTF samples were calculated on as is basis (mcg/100g). Each of the detected

folate compound was calculated as folic acid and sum of that provided total folate as folic acid and

reported as mcg/100g. Each sample was analyzed in three replicates on three separate days. The

averages and inter-day repeatability/precision (%RSD) calculated and are presented in the following

Table 17.

Table 17: Results of folate analysis in the SPIFAN II samples using the protease and amylase for

extraction but excluding the conjugase treatment. Total folate (mcg/100g) calculated as sum of

the detected folate compounds as folic acid in reconstitute beverages/RTF liquid samples.

Averages and precision (as % RSD) of replicates analyzed on separate 3 days (total n=9).

Sample

Matrix

Lot#

Analyzed

AS

Total Folate as Folic Acid

n

Repeat

days

#

Av. Mcg/100g

Liquid or RTF*

Stdev

% RSD

NIST SRM

1849a**

Infant Formula

-

Powder

210.6

3.3

1.6

3

1

BCR 485** Dried Vegetable

Powder

-

Powder

259.2

11.4

4.4

3

1

NIST SRM

3233**

Breakfast

Cereal

-

Powder

1233.6

41.1

3.3

3

1

LCS**

Breakfast

Cereal

-

Powder

39.3

4.4

11.1

5

3

2011.06/Fol-22 (February 2016)

FOR ERP USE ONLY

DO NOT DISTRIBUTE