ESTRO IMRT course 2015

Dose Calculation in Static and Rotational IMRT

Brussels October 8th, 2015

-57-

Dose calculation issues ‚in practice‘:

Dosimetric effects of the treatment table/setup devices

Reiner et al., DEGRO 2013

measured dose difference

with vs. without table [%]

gantry angle

15MV

6MV,

different

depths in phantom

•

depending on the vendor/model,

treatment tables can have a water-

equivalent thickness of ~1 cm w.r.t

dose absorption

•

depending on beam angle, dose

absorption of several percent

→

should be explicitly considered

for dose calculation, especially for

rotational techniques (VMAT),

where oblique angles through the

table cannot be avoided

•

Methods: Dose calculation with

table/setup devices ‘as in CT’, or

dummy contour with water density

override. MIND: evaluate optimal

method for your TPS/dose

calculation algorithm!

solid water phantom

with ionisation chamber

10x10 cm field,

different beam angles