Background Image
Table of Contents Table of Contents
Previous Page  1327 / 1544 Next Page
Information
Show Menu
Previous Page 1327 / 1544 Next Page
Page Background

ESTRO IMRT course 2015

Dose Calculation in Static and Rotational IMRT

Brussels October 8th, 2015

-57-

Dose calculation issues ‚in practice‘:

Dosimetric effects of the treatment table/setup devices

Reiner et al., DEGRO 2013

measured dose difference

with vs. without table [%]

gantry angle

15MV

6MV,

different

depths in phantom

depending on the vendor/model,

treatment tables can have a water-

equivalent thickness of ~1 cm w.r.t

dose absorption

depending on beam angle, dose

absorption of several percent

should be explicitly considered

for dose calculation, especially for

rotational techniques (VMAT),

where oblique angles through the

table cannot be avoided

Methods: Dose calculation with

table/setup devices ‘as in CT’, or

dummy contour with water density

override. MIND: evaluate optimal

method for your TPS/dose

calculation algorithm!

solid water phantom

with ionisation chamber

10x10 cm field,

different beam angles